          Operating instructions: Alarm

1.

Acknowledge the alarm.

2.

Check whether the alarm indicates high or low pressure. If the 

indicated pressure

 is correct, 

can 

be a malfunction to

 the alarm system or the pressure has returned to normal operating pressure.

3.

If the pressure is not correct: Check 

if

 the inlet pressure is correct. Change the operating side.

 

Fault trace according to local 

routines.

Operation

 and Maintenance

The 

line

 pressure regulator must be exercised annually and tested for safety. If no abnormal inertia, 

leakage or malfunction is detected, the device is approved for another year. In the event of a leak, 

replace the O-rings at the connection or valve body, after which a new safety test takes place. Safety 

testing takes place in accordance with national standard SIS HB 370.

Note that oil, grease and other contaminants must not come into contact with the product.

 

Regulators and safety valves can be serviced or replaced without interruption, by isolating the circuit to 

be serviced with the shut-off valves. Note that during service during operation, a circuit must always be 

kept open.

At annual safety test:

• The safety valves are checked to open at the set pressure. Also test the valve by pulling the relief

 

device on the side of the valve.

• The 

pressure gauge

 must be checked to show the correct pressure.

• O-rings are replaced if necessary.

• The function of the gas outlets must be tested.

After maintenance and service, the system must be leak tested. This is done by 

pressurizing

 the 

system and then checking that the system is 

leak 

tight with the help of leak spray.

Settings, repairs and service must be performed by Qmt-tech technicians or by technicians authorized 

by this company.

At product end, the unit is

 returned to an authorized recycling company.

6

When the alarm sounds

, take the following steps:

 rises above or decreases below set thresholds. 

The pressure transducers sends a signal to the alarm unit which sounds an alarm when the pressure

Summary of Contents for QMT 7MS028 Series

Page 1: ...ine pressure regulator 210331 3 Art nr QMT 7MS028XXX QMT 7MS035XXX QMT 7MS042XXX QMT 7MS054XXX Phone 46 0 480 44 02 00 Telefax 46 0 480 44 00 10 Address Amerikav gen 6 393 56 KALMAR Sweden Website e p...

Page 2: ...Contents 3 4 4 4 5 5 6 6 7 8 8 8 2 instructions...

Page 3: ...ly accessible if necessary more copies can be ordered from Qmt tech If something is unclear you can also call us at 0480 440200 Ensure the following The medical gases must not come into contact with o...

Page 4: ...outlet nominal distribution pressure is standard Pressure transducer for inlet nominal supply system pressure is option Optionally contact gauges can be fitted instead of pressure transducers The uni...

Page 5: ...lockwise Check against the pilot regulator pressure gauge and the outlet pressure gauge Lower the operating pressure by carefully opening the adjusting screw C on the regulator outlet side until a low...

Page 6: ...ators and safety valves can be serviced or replaced without interruption by isolating the circuit to be serviced with the shut off valves Note that during service during operation a circuit must alway...

Page 7: ...70 and the hospital s instructions The components of the brazing pressure regulator must not be exposed to temperatures exceeding 100 C either during installation or operation Should this happen or su...

Page 8: ...4 Black 2 Blue Connect to low alarm Connect to common conductor Connect to high alarm Spare parts QMT 78500XX QMT 7611XXX QMT 7200013 QMT QMT article number for Reulators Valves Pressure gauges Transm...

Reviews: