QMTECH QMT 7MS028 Series Operating & Maintenance Manual Download Page 4

Line

 pressure regulator principle

Qmt-tech's 

line

 pressure regulator 

(pressure stabilizing unit)

 is intended to regulate the pressure of 

medical gases from 

nominal supply system pressure to nominal distribution pressure.

 According to the 

principle of 

single fault condition

*, the unit contains two mutually independent operating circuits. One is 

intended for normal operation and the other as a reserve during, for example, service and inspection. 

This means that it is possible to replace regulators and safety valves during operation. 

Pressure 

transducer for outlet (nominal distribution) pressure is standard. Pressure transducer for inlet (nominal 

supply system) pressure is option. Optionally contact gauges can be fitted instead of pressure 

transducers.

 

The unit is equipped with safety relief valves on the outlet pressure side for each circuit.

 

These are equipped with a common 

relief gas

 pipe. The gas outlets are used as a measuring point.

*

 

Single fault condition:

 Situation in which a single component, which is to protect against a safety risk 

in an

 

equipment, is defective, or when a single external, abnormal circumstance is present.

 Flow

 

data

 Flow

 

chart 

Line

 pressure regulator with 

pressure sensor

P

I

Valve

Pressure regulator

Safety relief valve

Pressure gauge

Pressure 

transducer

 

Gas outlet

P

I

*

)

 

Optional

*

4

Summary of Contents for QMT 7MS028 Series

Page 1: ...ine pressure regulator 210331 3 Art nr QMT 7MS028XXX QMT 7MS035XXX QMT 7MS042XXX QMT 7MS054XXX Phone 46 0 480 44 02 00 Telefax 46 0 480 44 00 10 Address Amerikav gen 6 393 56 KALMAR Sweden Website e p...

Page 2: ...Contents 3 4 4 4 5 5 6 6 7 8 8 8 2 instructions...

Page 3: ...ly accessible if necessary more copies can be ordered from Qmt tech If something is unclear you can also call us at 0480 440200 Ensure the following The medical gases must not come into contact with o...

Page 4: ...outlet nominal distribution pressure is standard Pressure transducer for inlet nominal supply system pressure is option Optionally contact gauges can be fitted instead of pressure transducers The uni...

Page 5: ...lockwise Check against the pilot regulator pressure gauge and the outlet pressure gauge Lower the operating pressure by carefully opening the adjusting screw C on the regulator outlet side until a low...

Page 6: ...ators and safety valves can be serviced or replaced without interruption by isolating the circuit to be serviced with the shut off valves Note that during service during operation a circuit must alway...

Page 7: ...70 and the hospital s instructions The components of the brazing pressure regulator must not be exposed to temperatures exceeding 100 C either during installation or operation Should this happen or su...

Page 8: ...4 Black 2 Blue Connect to low alarm Connect to common conductor Connect to high alarm Spare parts QMT 78500XX QMT 7611XXX QMT 7200013 QMT QMT article number for Reulators Valves Pressure gauges Transm...

Reviews: