6. USING THE
QMI 1000
AS A THERMOMETER OR
PRESSURE METER
With the
QMI 1000
switched off, press and hold down the
/
button and then press and release
/
. Release
/
after
MANO_MOD
is displayed on top line.
The
QMI 1000
will now operate as a fixed display thermometer/pressure meter
with the pump off and inhibited.
The display will show:
P
0.00
→
Real time pressure reading.
T1
21.3
→
Use the T1 connection for the flow
temperature sensor.
T2
21.3
→
Use the T2 connection for the return
temperature sensor.
∆
T
0.0
→
Real time temperature difference.
The rotor display indications will now be locked apart from MENU. Readings
can be printed but not stored. Exit this mode by switching the
QMI 1000
off.
The standard printout for this mode is as follows:
K455 1.0
YOUR COMPANY NAME &
PHONE NUMBER HERE
DATE 15/05/07
TIME 13:00:47
…………….…………………………
T1
o
C 21.3
T2
o
C 21.3
∆
T
o
C 0.0
PRS mbar
0.00
………………………………………
Ref.
………………………………………