background image

Troubleshooting  Guide 

Error codes  displayed  Ьу the Tissuelyser  11 

"FOl" displayed 

Drive  motor of Tissuelyser  11 

overloaded 

"F02"  displayed 

Drive  motor of Тissuelyser  11 not 

starting 

"FОЗ"  displayed 

Specified  oscillation  frequency 

exceeded 

"F04"  displayed 

Transparent  hood  not  lowered 

"F05"  displayed 

Keys  held  down  fог too  long 

"F06"  displayed 

Speed  sensor defective 

"F07"  displayed 

Solenoid  switch  defective 

Tissuelyser  11 User  Manual  05/2016 

Comments and suggestions 

When the  drive  motor is  overloaded,  the  Tissuelyser  11 switches  itself off to  allow 
the  drive  motor to  сооІ.  А  display on  the  Тissuelyser  11 alternates  between  "FOl" 
and  the  remaining  cooling  time  in  minules  (e.g.,  "03" indicates З minutes). 
After the  cooling  time  has  elapsed,  switch  the  TissueLyser  11 off and  then  оп 
again  Ьеfоге  continuing  to  operate the  instrument. 

lf the  drive  motor does  not slart after the  "START"  key  has  been  pressed,  contact 
QIAGEN Technical  Services  Іо  request instrument  гераіг.  The  drive  motor  may 
Ье  blocked,  ог the  motor ог speed  sensor  may Ье  defective. 

lf the  specified  oscillation  frequency is  exceeded,  the Tissuelyser  11 switches  itself 
off and  "FОЗ"  is  displayed.  Contact QIAGEN Technical  Services  !о  request 
instrument  гераіг.  The  speed  sensor  may Ье  defective. 

То  protect the  operator,  the  drive  motor of the  Тissuelyser  11 will only start  if the 
transparent hood  has  been  lowered.  lf the  "START"  key  is  pressed  while the 

hood  is  raised,  "F04" is  displayed  and  the  drive  motor does  по! slart.  Іf the 
hood  is  raised  while the  Tissuelyser  11 is  in  operation,  the  drive  motor switches 
itself off and  "F04"  is  displayed.  То  cancel  the  "F04" display,  press  the  "STOP" 
key. 

Іf  one ог  mоге  keys  оп  the  keypad  аге  pressed  and  held  down  fог  mоге  than 

15  seconds,  the  drive  motor of the  Tissuelyser  11 switches  itself off and  "F05"  is 

displayed.  This  is  а safety  mechanism  to  prevent unwanted  changes  іп  the 
operation  of the  Тissuelyser  II  should  оп  object land  оп  the  keypad.  То  cancel 
the  "F05" display,  switch  the Tissuelyser  11 off and  then  оп  again. 

lf one of the  magnets  сопІгоІІіпg  the  speed  of the Tissuelyser  11 is  missing  ог 
defective,  the  instrument switches  ilself off and  "F06" is  displayed.  Contact 
QIAGEN Technical  Services  !о  request instrument  гераіг. 

Іf  there  is  а defect in  the  саЬІе  fог the  solenoid  switch,  which  monitors  the 
raising  of the  transparent hood,  the  Тissuelyser  11 switches  itself off and  "F07" is 
displayed.  Contact QIAGEN Technical  Services  to  request  instrument  героіг. 

Reviews: