QIAcube Connect MDx User Manual 12/2020
90
7.
On the data exchange screen, you can also create a download package, which contains
additional information like the logfiles. You can also check the remaining disk space.
Important
: If you use
Delete Files
, ensure to store the files from the USB stick at a safe place.
5.8
Low space on flash memory
If the flash memory disk is full, you cannot start a run. The system will show an alert and guide you
to the download screen. After downloading the files, confirm if they are successfully saved to the
USB stick, then store the files properly. When the backup is confirmed, you can delete the files on
the QIAcube Connect MDx USB flash drive.
Full disk space system alert
5.9
Independent heater/shaker operation
The heater/shaker can be operated individually if the QIAcube Connect MDx is not running a
protocol. The heating and shaking functions are not interlinked and can be used independently or
in combination.
Do not attempt to move the QIAcube Connect MDx during operation.
WARNING
Hot surface
[W21]
The shaker can reach temperatures of up to 70°C (158°F). Avoid touching it
when it is hot. Carefully remove the samples after a run.