Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
21
3. Press the timer up/down arrows until you reach the desired time remaining.
4. Press the shaking on/off button. The shaking indicator light will blink rapidly until the
desired speed is reached, then it will remain illuminated.
5. After the shaking indicator light is solidly illuminated, Press the timer on/off button to start
the countdown.
6. When the time display reaches zero (0:00), both the time and shaking functions will shut
off automatically. Four audible beeps will indicate the countdown function is complete,
and the time display will default back to the set time.
7. To interrupt an automatic timing cycle before it is completed, press the timer on/off
button. The time display will flash until you resume the time function by pressing the timer
on/off button again. This interrupt will not stop the shaking function, the shaking function
will stop only when the timer reaches zero (0:00).
5.4
Timer function – Accumulated time
1. By default, the timer will begin at zero (0:00) minutes. Press the timer on/off button to
begin timing.
2. Press the timer on/off button to stop timing. Press the timer on/off button again to
resume timing.
3. To reset the time to zero (0:00) minutes, make sure the timing is stopped, and press and
hold the timer on/off button for 3 seconds. Alternately, while of the timing is stopped,
simultaneously press the timer up and down arrows to reset the time to zero (0:00)
minutes.
5.5
Timer function – Remaining time
Note: If the timer is used in conjunction with the shaking function, when the time display
reaches zero (0:00), both the time and shaking functions will shut off automatically.
1. Press the timer up/down arrows until you reach the desired time remaining.