Qiagen Hybrid Capture Rotary Shaker 1 User Manual Download Page 1

Sample to Insight

__

 

September 2018 

Hybrid Capture

®

 System  

Rotary Shaker 1 User Manual 

6000-2110E (120 V) 
6000-2240E (230 V) 

QIAGEN 
19300 Germantown Road 
Germantown, MD 20874 

USA 

QIAGEN GmbH 
QIAGEN Strasse 1 
40724 Hilden 
GERMANY 

1108562   Rev. 01 

Summary of Contents for Hybrid Capture Rotary Shaker 1

Page 1: ...__ September 2018 Hybrid Capture System Rotary Shaker 1 User Manual 6000 2110E 120 V 6000 2240E 230 V QIAGEN 19300 Germantown Road Germantown MD 20874 USA QIAGEN GmbH QIAGEN Strasse 1 40724 Hilden GERMANY 1108562 Rev 01 ...

Page 2: ...09 2018 Trademarks QIAGEN Sample to Insight digene HC2 Hybrid Capture QIAGEN Group Registered names trademarks etc used in this document even when not specifically marked as such are not to be considered unprotected by law 2018 QIAGEN all rights reserved ...

Page 3: ...mation 6 2 1 Proper use 7 2 2 Electrical safety 9 2 3 Biological safety 9 2 4 Waste disposal 11 2 5 Symbols 12 3 Installation 14 3 1 Unpacking 14 3 2 Start up 14 4 Functional Description 16 4 1 Operational modes 17 5 General Operation 19 5 1 Loading Microplates 19 5 2 Continuous operation 20 5 3 Timer terminated operation 20 5 4 Timer function Accumulated time 21 5 5 Timer function Remaining time ...

Page 4: ...ng and replacing the shaker platform 25 6 4 Replacing the fuse 26 6 5 Shaking speed verification 27 6 6 Servicing 28 7 Troubleshooting 29 8 Technical Data 33 8 1 Operating conditions 33 8 2 Transport conditions 35 8 3 Storage conditions 35 Appendix A Waste Electrical and Electronic Equipment WEEE 36 Appendix B Warranty 37 Appendix C FCC Declaration 38 Ordering Information 40 ...

Page 5: ...tance For technical assistance and more information please see our Technical Support Center at www qiagen com TechSupportCenter or contact QIAGEN Technical Services or a local distributor 1 1 2 Version management This document is Hybrid Capture System Rotary Shaker 1 User Manual see the front cover of this user manual for document number and revision 1 2 Intended use The HCS Rotary Shaker 1 is int...

Page 6: ...tances are provided to avoid personal injury to you or other persons CAUTION The term CAUTION is used to inform you about situations that could result in damage to the instrument or other equipment Details about these circumstances are provided to avoid damage to the instrument or other equipment Before using the instrument it is essential to read this manual carefully and to pay particular attent...

Page 7: ...secure the HCS Rotary Shaker 1 to a level smooth stable surface by firmly pressing down on the four 4 corners of the unit creating a strong suction to the work surface DO NOT place on a bench mat Failure to do so will cause excess vibration and could cause personal injury Instrument damage and or property damage Risk of lnstrument damage Spills should be removed promptly DO NOT use a cleaning agen...

Page 8: ... device in close proximity to sources of strong electromagnetic radiation e g unshielded intentional RF sources as these can interfere with the proper operation Take the following precautions when operating or working near the HCS Rotary Shaker 1 Always make certain that the shaker platform and any hardware are secure before operating the HCS Rotary Shaker 1 Load the HCS Rotary Shaker 1 symmetrica...

Page 9: ...old room as soon as operation is complete 2 2 Electrical safety Only operate the HCS Rotary Shaker 1 with the power cord provided with the instrument For satisfactory and safe operation of the HCS Rotary Shaker 1 it is essential that the line power cord is connected to true electrical earth ground 2 3 Biological safety WARNING Hazardous substances The products used with this instrument may contain...

Page 10: ...cal analysis a potential biohazard that requires decontamination before reuse WARNING Risk of personal injury Sodium hypochlorite solution is caustic wear rubber gloves and eye protection when handling it To dispose of the HCS Rotary Shaker 1 follow all national state and local health and safety regulation and laws for disposing of laboratory waste For disposal of Waste Electrical and Electronic E...

Page 11: ...al 09 2018 11 2 4 Waste disposal Waste may contain certain hazardous chemicals or contagious biohazardous materials and must be collected and disposed of properly in accordance with all national state and local health and safety regulations and laws ...

Page 12: ...eral warning sign Type plate on the instrument CE mark for Europe Type plate on the instrument In vitro diagnostic medical device Type plate on the instrument RoHS mark for China the restriction of the use of certain hazardous substances in electrical and electronic equipment Type plate on the instrument Waste Electrical and Electronic Equipment WEEE SN Type plate on the instrument Serial number T...

Page 13: ...e instrument complies with applicable standards for electrical safety of laboratory equipment UDI Label on the instrument Global trade item number Instrument box label Fragile handle with care Instrument box label Consult instructions for use Front cover of this user manual Authorized representative in the European Community Front cover of this user manual Catalog number ...

Page 14: ...nstrument is operated successfully 3 2 Start up Make sure to select the correct voltage rating by checking the type plate on the side of the instrument Record the serial number located on the type plate for future reference Turn the unit upside down gently rest it on the shaking platform and inspect the suction cup feet Remove any dust or debris from the suction cup feet with 70 Isopropyl alcohol ...

Page 15: ...ench mat WARNING CAUTION Risk of personal injury and lnstrument damage Failure to create a strong suction to secure the Shaker will cause excess vibration personal injury Instrument damage and or property damage At a corner of the instrument attempt to push the instrument sideways with a medium amount of force If the instrument is properly suctioned to the surface it will not move Plug the power c...

Page 16: ...S Rotary Shaker 1 can be used in environments from 10 to 60 C allowing for cold room use as well as incubated applications The HCS Rotary Shaker 1 is constructed of heavy gauge metal that provides a stable base for a steady vibration free operation The following figures show the major external components of the instrument All of the operating controls for the HCS Rotary Shaker 1 are located on the...

Page 17: ...wn arrows The power inlet and fuse drawer are on the back of the HCS Rotary Shaker 1 Power inlet module Fuse drawer 4 1 Operational modes The speed of the Shaker 1 is variable from 100 1200 revolutions per minute RPM It operates in a circular motion with an orbit of 0 3 cm 0 12 inches The HCS Rotary Shaker 1 features a timer with an operating time range from 0 to 9 999 minutes in one 1 second incr...

Page 18: ...18 Hybrid Capture System Rotary Shaker 1 User Manual 09 2018 The HCS Rotary Shaker 1 can operate in both Continuous Operation or Timer Terminated operation modes ...

Page 19: ...haker 1 is designed to hold two 2 or four 4 microplates but must be loaded symmetrically to avoid unbalanced loads When shaking a single microplate a second empty microplate must be placed diagonally to balance the load Likewise when shaking 3 microplates one empty microplate must be placed in the fourth position to balance the load 1 Place two 2 microplates diagonally on the platform or place fou...

Page 20: ...ing press the shaking on off button The shaking indicator light will blink rapidly until the desired speed is reached then it will remain illuminated The HCS Rotary Shaker 1 will shake continuously until the shaking on off button is pressed 4 To stop the shaking press the shaking on off button 5 3 Timer terminated operation Timer terminated operation allows for timed shaking procedures 1 Power ON ...

Page 21: ...ume the time function by pressing the timer on off button again This interrupt will not stop the shaking function the shaking function will stop only when the timer reaches zero 0 00 5 4 Timer function Accumulated time 1 By default the timer will begin at zero 0 00 minutes Press the timer on off button to begin timing 2 Press the timer on off button to stop timing Press the timer on off button aga...

Page 22: ...on again 5 To interrupt an automatic timing cycle before it is completed press the on off button to the right of the time display The time display will flash until you resume the time function by pressing the on off button again This interrupt will not stop the shaking function the shaking function will stop only when the timer reaches zero 0 00 5 6 Beeper Preference 1 To silence beeper operation ...

Page 23: ...y Shaker 1 contact QIAGEN Technical Services QIAGEN charges for repairs that are required due to incorrect maintenance 6 1 Cleaning and decontamination WARNING CAUTION Risk of personal injury and material damage Consider any laboratory equipment used for research or clinical analysis a potential biohazard that requires decontamination before reuse The user is responsible for decontamination to the...

Page 24: ...wetted with a solution of 0 5 sodium hypochlorite solution NaOCl or bleach Industrial bleach contains approximately 10 NaOCl household bleach contains approximately 5 NaOCl When using industrial bleach prepare a 1 20 mixture of bleach to water When using household bleach prepare a 1 10 mixture of bleach to water Follow by wiping with a soft cloth moistened with deionized or distilled water 6 2 Reg...

Page 25: ... and replace the shaker platform as follows 1 Remove the 4 screws on the shaker platform that are covered when plates are loaded on the platform 2 Lift the shaker platform up off the shaker base 3 Clean the shaker platform and shaker base with a dilute detergent solution Make sure the shaker platform is completely dry before proceeding to the next step 4 Line up the 4 screw holes on the shaker bas...

Page 26: ...tly pry open the fuse drawer at the rear of the shaker base 3 Remove the fuse from the plastic clip in the fuse drawer 4 Place a new fuse in plastic clip in the fuse drawer 5 Insert the fuse drawer back into the instrument Voltage HCS Rotary Shaker 1 catalog number Fuse Amperage Fuse Type 120 V 6000 2110E 5 AMP 250 volt 5 x 20 mm 0 25 x 1 25 in Quick acting UL listed fuse 230 V 6000 2240E 5 AMP 25...

Page 27: ...t will blink rapidly until the desired speed is reached then it will remain illuminated 7 Press and hold the Start Measurement button on the tachometer Depending on the tachometer the functional description of the tachometer may vary 8 Aim the light beam onto the reflective tape so the reflective tape crosses the light beam once every revolution Focus the light beam on the reflective tape by raisi...

Page 28: ...ur local QIAGEN representative and provide full failure details When making your call please make sure that you have the serial number of the instrument Do not ship the instrument back for repair until advised to do so by your local representative or QIAGEN Technical Services In the event that you are requested to return the instrument or any part thereof it is your legal requirement to ensure tha...

Page 29: ...oning Make sure the power source has power correct as necessary Fuse may need to be replaced Replace the fuse See Replacing the fuse page 26 Speed and Time displays are not illuminated Power cord is not plugged in properly Make sure that the power cord is plugged in Power source is not functioning Make sure the power source has power correct as necessary Fuse may need to be replaced Replace the fu...

Page 30: ...te the instrument to a flat even surface Suction foot is loose Secure each suction foot to the benchtop by firmly pressing on all four corners of the unit If this does not resolve the issue clean all six suction feet and the benchtop with 70 isopropyl alcohol and a lint free cloth then secure the unit to the benchtop by firmly pressing on all four corners of the unit The shaker platform is loose S...

Page 31: ... of a letter V Tighten platform screws Remove foreign object and restart unit Rattling or ticking sound when shaking Loose screw on the platform Foreign object on the platform Error Code E04 is displayed unit overload Maximum load exceeded Remove excessive load from the platform Press the Power Standby button to clear this error Press the Power Standby button again to resume operation Loose suctio...

Page 32: ...andby button to clear this error Press the Power Standby button again to resume operation Loose suction foot Secure each suction foot to the benchtop by firmly pressing on all four corners of the unit Press the Power Standby button to clear this error Press the Power Standby button again to resume operation Drive system failure If the E03 error persists after the above troubleshooting steps contac...

Page 33: ... 4 inches Shipping Weight 11 4 kg 25 lbs Power requirements for 6000 2110E 120 Volts AC 50 60Hz Power requirements for 6000 2240E 230 Volts AC 50 60Hz Power consumption both voltages 20 Watts Fuse requirements both voltages 5 AMP 250V Quick acting Timer 0 to 9 999 minutes in one 1 second increments Air temperature 10 to 60 C Relative humidity Maximum 80 non condensing ...

Page 34: ...Manual 09 2018 Condition Parameter Maximum load 4 microplates Place of operation For indoor use only Pollution level II Altitude Up to 2000 meters 6562 feet Shaking speed 100 1200 RPM Shaking motion Orbital Shaking orbit diameter 0 3 cm 0 12 inches ...

Page 35: ...port conditions Condition Parameter Air temperature 20 to 65 C in manufacturer s package Relative humidity Maximum 80 non condensing 8 3 Storage conditions Condition Parameter Air temperature 20 to 65 C in manufacturer s package Relative humidity Maximum 80 non condensing ...

Page 36: ...oint for recycling according to local laws and regulations Separate collection and recycling of waste electronic equipment at the time of disposal helps to conserve natural resources and make sure that the product is recycled in a manner that protects human health and the environment QIAGEN provides recycling upon request at additional cost To recycle electronic equipment contact your local QIAGEN...

Page 37: ...anufacturer If notified of such defects during the warranty period the manufacturer will at its option either repair or replace products that prove to be defective The warranty shall not apply to defects resulting from improper or inadequate maintenance by the customer unauthorized modification or service misuse operation outside of the environmental specifications for the product or units returne...

Page 38: ...ul interference and 2 this device must accept any interference received including interference that may cause undesired operation This equipment has been designed and tested to CISPR 11 Class A In a domestic environment it may cause radio interference in which case you may need to take measures to mitigate the interference This Class A digital apparatus complies with Canadian ICES 0003 The followi...

Page 39: ...ouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio T V technician for help QIAGEN is not responsible for any radio te...

Page 40: ...nformation Product Contents Cat no Hybrid Capture System Rotary Shaker 1 120 volt rotary shaker for use with digene Hybrid Capture 2 DNA tests 6000 2110E Hybrid Capture System Rotary Shaker 1 230 volt rotary shaker for use with digene Hybrid Capture 2 DNA tests 6000 2240E ...

Page 41: ...Hybrid Capture System Rotary Shaker 1 User Manual 09 2018 41 This page intentionally left blank ...

Page 42: ...42 Hybrid Capture System Rotary Shaker 1 User Manual 09 2018 This page intentionally left blank ...

Page 43: ...Hybrid Capture System Rotary Shaker 1 User Manual 09 2018 43 This page intentionally left blank ...

Page 44: ...44 Hybrid Capture System Rotary Shaker 1 User Manual 09 2018 Ordering www qiagen com contact Technical Support support qiagen com Website www qiagen com ...