QHY16200A
Stretch to background level
Stretch to Max Range
A typical 10sec 4*4binning Dark Frame with all the region of the image. The vertical blank
banding on the right of the image is the overscan area. The overscan area is important for
overscan calibration. We recommend you reserve this area. You can also see two strip that is a
little brighter than the image area. It is the "Optic Black" Area.
The difference of "Optic Black" Area and "Overscan" Area is that the "Optic Black" area has a
metal shield on it and it is not sensitive to light, But thermal current still exists. The "Overscan"
area is a virtual area and it represents the bias.