6
TROUBLESHOOTING
Troubleshooting
No sound when unit is turned on.
1.
Check Battery and circuit board by following Step 1 of Equipment Check, above. Replace
battery if low and make sure wire connections of panel are tight.
2.
Repeat Step 2 of Equipment Check, above. Check probe conductor to make sure it is clean
and not encrusted with mineral deposits. Check tape/probe connection and tape for any
breaks.
Continuous sound when unit is turned on or removed from water.
1.
Make sure MP30 has "Drawdown Control" switched to "OFF" position.
2.
Make sure probe co
nductor tip is clean.
3.
Check for excess moisture on the back of the electronic panel and wire connector area.
4.
Check probe/tape connection and tape for any breaks or leaks where water might get in.