background image

Manual

Part No.  95180    5-31-07  

MP30

Draw Down Meter

MODEL

PATENT PENDING

O

U

asics

Instruction

P.O. Box 3726   Ann Arbor, MI  48106-3726  USA

1-800-624-2026    Fax (734) 995-1170

[email protected]    www.qedenv.com   

Summary of Contents for MP30

Page 1: ...Manual Part No 95180 5 31 07 MP30 Draw Down Meter MODEL PATENT PENDING O U asics Instruction P O Box 3726 Ann Arbor MI 48106 3726 USA 1 800 624 2026 Fax 734 995 1170 info qedenv com www qedenv com ...

Page 2: ......

Page 3: ...ce Contacts Cleaning the Meter Page 1 1 1 2 3 5 5 5 3 4 4 6 7 7 8 MP 30 Operating Instructions Background Drawdown Control Quick Guide Drawdown Control Response with MicroPurge basics controllers General Information on MP30 Features and Usage Use in the Field Static Water Level Measurement Troubleshooting QED Warranty ...

Page 4: ...s 3 When switched to water level mode the MP30 provides conventional water level meter functions of light and sound indication upon contact of the probe with water Drawdown Control Quick Guide 1 Determine static water level in conventional manner with MP30 power switched ON and Drawdown Control switched to OFF see Figure 1 2 When well purging is to begin switch the MP30 Drawdown Control switch to ...

Page 5: ...e probe to a new lower drawdown control point to increase the well recovery rate Consult regulatory considerations before adjusting protocols 2 Figure 1 Speaker Unit On Off Switch Panel Knobs On Off Switch Indicator Level Shut Off Indicator Contol Port MP30 Cable connection 1 Cable connects to identical ports on MP30 and basics controllers labeled Control Port 2 Connect cable to MP30 after reel ha...

Page 6: ...e Micropurge basics controller for detailed information In general when the Drawdown Control mode is switched ON and the probe becomes dry the controller display will indicate LVL in the upper left hand corner of the controller display and the con troller will pause pump operation until the probe again becomes submerged When the probe is again submerged the basics controller will eliminate the LVL...

Page 7: ...ot respond make sure the panel securing knobs are tight see figure 1 If the unit still does not respond replace the 9 volt battery reached by removing the securing knobs and panel 2 Test tape and probe by shorting out the center conductor and probe body on the pin on the back axle of the unit while touching the white band on the probe with a moistened finger or lower the probe into a container of ...

Page 8: ...ontamination carryover 2 Unwind the tape and wash with mild detergent then rinse well and rewind The following cleaning agents are acceptable 10 Alconox Joy or Lestoil methyl isopropyl and isobutyl alcohols hexane and heptane Rinse thoroughly with water afterwards 3 Wash reel if necessary The central electronic panel can be removed and the reel washed down Unthread the securing knobs disconnect th...

Page 9: ...e conductor to make sure it is clean and not encrusted with mineral deposits Check tape probe connection and tape for any breaks Continuous sound when unit is turned on or removed from water 1 Make sure MP30 has Drawdown Control switched to OFF position 2 Make sure probe conductor tip is clean 3 Check for excess moisture on the back of the electronic panel and wire connector area 4 Check probe tap...

Page 10: ...e Hot Line 1 800 624 2026 1 734 995 2547 1 734 995 1 170 1 800 272 9559 service qedenv com MP30 14 x 10 1 2 x 8 37 x 27 x 20 cm 7 lb 3 2 kg w 150 tape 9 lb 4 kg w 300 tape 5 8 O D 1 6 cm 7 5 19 cm Optional Included Included Included 9 Volt Battery 30 40 Hours 150 or 300 Feet 46 or 91 m 40 to 185 F 40 to 85 C SPECIFICATIONS Weight Probe Diameter Probe Length Well Hanger Operating Temperature ...

Page 11: ...e defective product with a comparable new unit at no charge to the Buyer or to refund the full purchase price In no event shall such allegedly defective products be returned to QED without its consent and QED s obligations of repair replacement or refund are conditioned upon the Buyer s return of the defective product to QED IN NO EVENT SHALL QED ENVIRONMENTAL SYSTEMS INC BE LIABLE FOR CONSEQUENTI...

Page 12: ...age If the return of a component is determined to be necessary QED will authorize the return of the component at owner s expense If the product proves not be defective within the terms of this warranty then all costs and expenses in connection with the processing of the Purchaser s claim and all costs for repair parts and labor as authorized by owner hereunder shall be borne by the Purch aser RESP...

Page 13: ......

Page 14: ...P O Box 3726 Ann Arbor MI 48106 3726 USA 1 800 624 2026 Fax 734 995 1170 info qedenv com www qedenv com ...