background image

OPERATION

10

Sample Collection

  The PAUSE key (           ) is used to freeze the controller action to allow the 

user time to collect a sample or carry out other steps that might be difficult if the controller contin-
ued to automatically cycle and cause the pump to produce water.  While the controller is cycling, 
pressing the PAUSE key causes the controller to immediately enter the Hold state.  Drive air is 
vented from the pump (this is the pump refill cycle) and the pump fills and waits.  Pressing the 
PAUSE key a second time causes the controller to immediately enter the Sample state.  Drive air 
is directed to the pump causing the pump to discharge its volume of liquid.  Bladder pumps typic-
ally hold 400-500 ml of liquid, so use of the Hold and Sample states allow the full volume of the 
pump to be discharged into a sample container.  Pressing PAUSE once again returns the MP10 to 
its normal Automatic Cycling state.  During Hold and Sample a HELD is displayed to remind you 
that the controller is in a paused state.  Figure 3 shows an example of the MP10 in MP mode, but 
HELD in the Sample state.  

 

I I

Figure 3 MP10 Held State (MP mode)

MicroPurge Basics Controller

MP Micro Purge
ID   ID Time Set

MN   Manual Time Set
LVL  LevelShutoff

CYCLE

I I

MODE

CPM/Value

Flow/Value

Flow/Value

Hold/Sample

/Cycle

Start/Stop

ID/MN/MP/

Battery

Mode

CPM

ID

Time

Refill

Discharge

HELD 103 10.0

MPCPM4 >05.0

   D

   E

Each of the 165 possible ID settings corresponds to a unique ID that is associated with CPM, refill 
and discharge time values.  For typical usage, only the UP/DOWN arrow keys are required to set 
flow, and the ID number is provided for easy, direct return to past settings. Appendix 1 lists all pos-
sible ID settings and the default refill and discharge time settings for each CPM.  Appendix 1 also 
shows how the refill and dis-charge time will change within a CPM setting as you press the UP or 
DOWN keys.

Note: Pressing the Cycle key also freezes controller cycling. However, using the Cycle key rather 
than the Pause key causes the startup screen to be displayed upon restart. Use of the Pause key 
is recommended for typical operation.

Summary of Contents for Micropurge MP10

Page 1: ...Part No 95177 9 23 07 MP10 CONTROLLER MODEL PATENT PENDING O U asics Manual Instruction P O Box 3726 Ann Arbor MI 48106 3726 USA 1 800 624 2026 Fax 734 995 1170 info qedenv com www qedenv com...

Page 2: ...MP10 Controller Optional MP30 Drawdown Meter Well Cap Tubing Air Supply In Pump Air Supply Pump Discharge Well Wizard Bladder Pump R MP10 Basic Setup...

Page 3: ...icroPurge Mode Using CPM Sampling Use with the MP30 Level Sensing Water Level Meter ID Mode User Time Set MN Mode MP10 Battery Troubleshooting MP10 Specifications QED Service Contacts QED Warranty App...

Page 4: ...trol Battery Cover Flow throttle Discharge cycle indicator psi 0 THROTTLE CONTROL PORT DISCHARGE CYCLE AIR IN AIR OUT MicroPurge Basics Controller MP Micro Purge ID ID Time Set MN Manual Time Set LVL...

Page 5: ...Set MN Manual Time Set LVL LevelShutoff CYCLE I I MODE CPM Value Flow Value Flow Value Hold Sample Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00 1 10310 0 MPCPM4 05 0 D E CYCL...

Page 6: ...f pump is Refilling or Discharging Cycles per minute setting The mode the controller is in Refill time setting in seconds Time remaining in the cycle ID for time settings Abbreviations CPM MP ID HELD...

Page 7: ...each well ID Time Set Mode Operation Quickly recalls pre determined settings for each well by specifying a 3 digit ID Level Delay Interface The controller plugs into the optional MP30 MicroPurge Draw...

Page 8: ...Page 6 Follow instructions on the battery panel Opening the MP10 case turns power ON Note It may take up to 20 seconds for the unit to power up Select desired Cycles Per Minute CPM with default value...

Page 9: ...ate the refill time should be set long enough to achieve the maximum volume of air bubbles on each pump cycle and the discharge time should be set long enough to ensure that the air has stopped bubbli...

Page 10: ...stored with its lid closed Opening Display The opening display is shown for 5 seconds and displays the controller name the version number and the battery voltage as shown in Figure 1 Figure 1 shows t...

Page 11: ...D mode Figure 2 MP10 MicroPurge Mode Using CPM The MP10 introduces a revolutionary simpler way to control bladder pump flow rate and achieve the low flow method used by experts Up down arrow keys are...

Page 12: ...And the MP10 assigns a unique identification value to each setting the ID value which can be directly set during later sampling events MP displayed in the lower left corner of the display indicates Mi...

Page 13: ...o remind you that the controller is in a paused state Figure 3 shows an example of the MP10 in MP mode but HELD in the Sample state I I Figure 3 MP10 Held State MP mode MicroPurge Basics Controller MP...

Page 14: ...Drawdown Control The MP10 may optionally be used with the MP30 MicroPurge Drawdown Water Level Meter See Figure 4 for an example of the MP10 in MP mode with the controller in a level paused state ena...

Page 15: ...Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00 1 10310 0 MPCPM4 05 0 D E D E 1 800 624 2026 micropurge com O U ED asics CONTROLLER MP10 MicroPurge Mode Quick Guide 3 AA BATTER...

Page 16: ...ra tion so as to maintain drawdown at the set level When the water level drops below the probe the MP30 sends a signal to the MP10 to pause pumping Both the MP10 and the MP30 give visual signals and t...

Page 17: ...Set MN Manual Time Set LVL LevelShutoff CYCLE I I MODE CPM Value Flow Value Flow Value Hold Sample Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00 1 10310 0 MPCPM4 05 0 D E 14...

Page 18: ...YCLE I I MODE CPM Value Flow Value Flow Value Hold Sample Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00 1 10 0 MN 05 0 D E 15 Note changes in time settings that are entered wh...

Page 19: ...to select the digits of the refill and discharge time settings found at the rightmost positions on the display The UP and DOWN keys are used to adjust the digit value up or down By selecting and adjus...

Page 20: ...erged when in MP mode Temperature below 10F Warm controller Air not cycling through controller Throttle turned too low Turn throttle clock wise to produce pressure Air source not delivering air Verify...

Page 21: ...humidity Circuitry protected against transient surges introduced from improper battery installation or switch connect ions LCD display 32 character 2 lines 16 characters each Non glare double hardened...

Page 22: ...he defective product with a comparable new unit at no charge to the Buyer or to refund the full purchase price In no event shall such allegedly defective products be returned to QED without its consen...

Page 23: ...e If the return of a component is determined to be necessary QED will authorize the return of the component at owner s expense If the product proves not be defective within the terms of this warranty...

Page 24: ...13 5 13 12 5 12 11 5 11 10 5 9 5 9 8 5 8 7 5 7 6 5 6 5 5 5 10 30 31 32 33 34 35 36 37 38 39 40 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38...

Page 25: ...P O Box 3726 Ann Arbor MI 48106 3726 USA 1 800 624 2026 Fax 734 995 1170 info qedenv com www qedenv com...

Reviews: