Declaration – electromagnetic emissions and immunity for equipment and systems that are not life-supporting
and are specified for use only in a shielded location
The QardioCore Wireless ECG Monitor system declaration-electromagnetic immunity
The QardioCore Wireless ECG Monitor system is intended for use in the electromagnetic environment specified below.
The customer or the user of the QardioCore Wireless ECG Monitor system should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3V
Portable and mobile RF communications
equipment should be used no closer to
any part of the EQUIPMENT or SYSTEM
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency
of the transmitter. Interference may
occur in the vicinity of equipment
marked with the following symbol
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3V/m
385 MHz | 5 V/m
450 MHz | 10 V/m
810 to 930 MHz | 20 V/m