background image

INSERT OR REMOVE

INSTRUMENT

OPERATION

CLEANING & CARE

I N S T R U M E N T S

STEP 3:

 Insert or remove the 

instrument as needed.

STEP 4:

 Release the collar and it will 

snap back into position.

STEP 1:

 Grasp ACMI Quick Disconnect 

Collar between Thumb and fore fi nger 
as shown.

STEP 2:

 Pull collar back toward the black 

handle of the instrument adapter. It will 
only move about 3 mm.

PULL BACK

The fi ber optic light transmitting surfaces 
on both ends of the fi ber optic clip or wand 
should be wiped clean with isopropyl alcohol 
before autoclaving or chemiclaving.

In the event that a fi ber optic clip becomes 
loose, 

gentle

 pressure can be applied with 

pliers to tighten clips.

STERILIZATION

 

Your fi ber optic instrument may be autoclaved 
or chemiclaved. Do not exceed 275°F (135 C). 
Please note that exposure to heat will reduce 
the life of your fi ber optic instrument. The 
average life expectancy of fi ber optics is 18 to 
24 months with autoclaving or chemiclaving. 
This time could be longer or shorter 
depending on frequency and temperature of 
autoclave or chemiclave.

DO NOT attempt to sterilize the 

Instrument Cable or Battery Pack.

See reverse side for cleaning and care 

of instrument cable and battery pack.

On/Off button:

Turn on & off,
Comes on at last used 
intensity  setting

Increase intensity button:

10 positions, beeps when 
pressed, long beep at 
maximum intensity

Decrease intensity button:

10 positions, beeps when 
pressed, long beep at 
minimum intensity

Low Battery Warning:

5 short beeps, every 60 
seconds, starts with about 7 
minutes life remaining

Use circular motion

Connect instrument 

adapter cable to 

battery pack and 

enjoy your Q-Optics 

Radiant

 Light System.