background image

8.2

Bluetooth Connection to Devices

The SP-380 uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone
via the

uPyxis®

Mobile App or to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the

uPyxis®

Desktop App. To allow the SP-380 to connect via Bluetooth with other devices follow the steps

below:

1. Power on the SP-380 by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Diagnosis

(

) to launch the

SYSTEM DIAGNOSIS

screen.

4. Allow 5–10 seconds for the message in the top-right corner of the display change from

Starting BTLE...

to

BTLE Started

(Figure 13).

5. Choose to connect via one of two options:

(a) The

uPyxis®

Mobile App (see the

Use with uPyxis® Mobile App

section), or

(b) The

uPyxis®

Desktop App (see the

Use with uPyxis® Desktop App

section).

8.2.1

Calibrate an ST-500 Series Sensor with the SP-380 via Bluetooth

The SP-380 can be used to verify the result of an inline Pyxis ST-500 Series sensor by measuring the sample
water taken from the inline sensor sample line. The SP-380 can then be used to calibrate the inline sensors
over the Bluetooth connection. To calibrate an inline sensor, follow the steps below:

1. Power on the SP-380 by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Comm

(

) to launch the

COMMUNICATION

screen (Figure 17).

4. Press

Scan

(

) to begin scanning for Bluetooth devices.

5. Discoverable devices will begin to populate on the display with their name and MAC-Address (Figure

18).

6. If more than one device appears in the

Device list

, press

»

(

) to cycle through the devices.

7. If no devices or the incorrect device appear in the

Device list

, press

Scan

(

) to re-scan for discov-

erable devices.

8. Press

Connect

(

) to begin pairing to the selected sensor.

9. When the connection is established, the SP-380 displays the latest PTSA measurement from the con-

nected sensor (Figure 19).

10. Fill the main module sample cell with the same sample water that the sensor is measuring.

11. Press

Read

(

) to see PTSA measurement from the SP-380along with the sensor measurement

(Figure 20).

12. Press

Calib

(

) to begin sensor PTSA calibration.

SP-380 User Manual

[email protected]

| +1 (866) 203-8397

13

Summary of Contents for SP-380

Page 1: ...Images _manualcover jpg...

Page 2: ...dheld PTSA Fluorescein Dual Meter User Manual December 15 2020 Rev 1 31 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Con...

Page 3: ...8 Device Information and Diagnosis 11 8 1 Sample Cell Cleanliness Check 12 8 2 Bluetooth Connection to Devices 13 8 3 Factory Reset 15 9 Use with uPyxis Mobile App 16 9 1 Download uPyxis Mobile App 16...

Page 4: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Page 5: ...y interference Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 PTSA Specifications Item Specification Range 0 300 ppb Accuracy 1 or 1 0 ppb Resolution 0 1 ppb E...

Page 6: ...310 g Humidity 85 at 106 F 41 C Enclosure Rating IP67 With Pyxis s continuous improvement policy these specifications are subject to change without notice Batteries excluded 3 Unpacking Instrument Re...

Page 7: ...dheld Cleaning Kit SER 02 4 Installation 4 1 Battery Installation The SP 380 is powered by four alkaline batteries Typical battery life lasts for 10 000 measurements or 10 months When the battery capa...

Page 8: ...he SP 380 has three control keys as shown in Figure 2 The left right and ok keys are used to launch actions indicated on the LCD display directly above the keys The labels above the keys indicate the...

Page 9: ...The values will be displayed in white with a blue background if a stable value is reached Figure 2 For a sample containing 100 ppb PTSA the measured PTSA should be stabilized within the range of 98 10...

Page 10: ...nning a measurement use the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 380 to stabilize The values will be displayed in white with a blue background if a...

Page 11: ...art the zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 5 7 Rinse the sample cell three times with the desired PTSA stand...

Page 12: ...he DI water 2 Power on the SP 380 by pressing Allow 5 10 seconds for the SP 380 to stabilize 3 Press FC to switch to Fluorescein measurement mode 4 Press to launch the FLCN CALIBRATION screen Figure 7...

Page 13: ...nt calibration or b Press Exit to end the calibration process at a three point calibration and return to Fluo rescein measurement mode 7 2 3 Four Point Calibration 15 Rinse the main module sample cell...

Page 14: ...or main module also shown Press Diagnosis to launch the SYSTEM DIAGNOSIS screen where raw measurement data are displayed Figure 13 The information has no use for normal operation but instead is used f...

Page 15: ...INFORMATION screen 3 Press Diagnosis to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforthe messagein the top rightcornerof thedisplaychangefrom Starting BTLE to BTLE Started 5 Press Cleanline...

Page 16: ...ter taken from the inline sensor sample line The SP 380 can then be used to calibrate the inline sensors over the Bluetooth connection To calibrate an inline sensor follow the steps below 1 Power on t...

Page 17: ...ly the same as the value measured by the SP 380P ress Read again to take more readings from the sensor if nec essary 14 If the calibration is successful a Verify calibration OK message will appear on...

Page 18: ...to BTLE Started 5 Press Help to launch the HELP screen Figure 23 6 Press Factory Reset or The display updates as shown in Figure 24 and the user can choose one of three options a Press to start the f...

Page 19: ...le smart phone according to the following steps 1 Follow the steps in the Bluetooth Connection to Devices section to make the SP 380 discoverable 2 Open uPyxis Mobile App 3 On uPyxis Mobile App pull d...

Page 20: ...hen connected the uPyxis Mobile App will default to the Setting screen From the Setting screen the user can set the Power off time and Screen off time in seconds Figure 27 SP 380 User Manual service p...

Page 21: ...e press Get Firmware Once the new firmware is downloaded press Upgrade Firmware NOTE The firmware update process takes some time and will require the SP 380 to stay within range approximately 10 ft wi...

Page 22: ...not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis Setup exe file to i...

Page 23: ...ices section to make the SP 380 discoverable 2 Plug the Bluetooth USB adapter into a USB port in the computer 3 Launch uPyxis Desktop App 4 On uPyxis Desktop App click Device Connect via USB Bluetooth...

Page 24: ...ate firmware file contact service pyxis lab com for these firmware files and clicking Upgrade Firmware NOTE The firmware update process takes some time and will require the SP 380 to stay within range...

Page 25: ...month The SP 380 will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Read Datalog deleted by clicking Delete or exported by clicking Expo...

Page 26: ...dark will reset the timer and the screen will re light This press will not step along the calibration process the next press needed will still be required in the sequence Always rinse the unit with cl...

Page 27: ...re condition such as leaving the SP 380 inside an unattended automobile 12 Contact Us Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com Phone 1 866 203 8397 Email service...

Reviews: