GUIDANCE AND MANUFACTURER’S
DECLARATION - IMMUNITY
The Squid Go is intended for use in the electromagnetic environment
specified below. The customer or user of the Squid Go should ensure
that it is used in such an environment.
ESD
IEC 61000-4-2
+6kV Contact
+8kV Air
+6kV Contact
+8kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%.
EFT
IEC 61000-4-4
+2kV Mains
+1kV I/Os
+2kV Mains
+1kV I/Os
Mains power quality should
be that of typical
commercial or hospital
environment.
SURGE
IEC 61000-4-5
+1kV Differential
+2kV Common
+1kV Differential
+2kV Common
Mains power quality should
be that of a typical
commercial or hospital
environment.
VOLTAGE
DIPS/DROPOUTS
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Squid Go requires
continued operation during
power mains interruptions,
it is recommended that the
Squid Go be powered from
an uninterruptible power
supply or battery.
POWER
FREQUENCY
50/60Hz
MAGNETIC FIELD
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
31
Summary of Contents for Squid Go
Page 2: ...GO DEEP ...
Page 13: ... SHOULDER WRAP LEG WRAP WRIST WRAP KNEE WRAP 10 ...
Page 39: ...GO DEEP ...