Ionic® Purification System
Purigen Biosystems
FOR RESEARCH USE ONLY.
Not for use in diagnostic procedures.
42
Self Test
The
Self Test
runs a set of automated diagnostics that checks many of the instrument functions. Once complete, a green check
indicates the diagnostic tests passed and a red x indicates one or more diagnostic tests did not pass. If the self test results in a red x,
the results should be emailed to
for interpretation.
1.
From the Maintenance & Service menu, press
Self Test
.
2.
The Ionic system begins performing a self test. Progress is indicated by the percent progress wheel on the screen. When the self
test is complete, press the arrow button on the left-side of the touchscreen to return to the
Maintenance & Service
screen.
FIGURE 37:
Self Verification Test screen
3.
If the
Self Test
is not successful, follow the steps in
on the next page to save the system log file to a USB
flash drive and then email the log file to Purigen Biosystems support.