background image

How to set the zero and full scale range 

 
This  will  need  to  be  set  up  in  the  field  to  adjust  to  the  length  of  cable  between  the  oxygen 
monitor  and  the  remote  display  alarm  indicator.    To  perform  this  in  the  field,  first  enter  the 
password on your oxygen monitor and select the “Set 4-20mA loop” menu 
 
Push  the  joystick  right  to  select  “Set  4mA  Zero”  and  then 

push  the  joystick  right  one  more 

time to activate it.

 This sends a 4mA signal to the remote display.  On the back of the remote 

display  push  the  ZERO  button  for  3  seconds.  You’ll  see  the  display  reset  and  automatically 
adjust itself to the new zero.  If necessary, you can push the joystick up or down to fine tune the 
zero reading.  

NOTE:

 

Both AL1 and AL2 LED’s will also be illuminated.

 

     (To exit the Zero menu, push the Joystick left once) 
 
Push the joystick down once to select “Set 20mA Span” and  
then 

push the joystick right one more time to activate it.

 This  

sends  a 20mA signal to the remote display. On the back of the  
remote display push the SPAN button for 3 seconds. You’ll see the  
display reset and automatically adjust itself to the new span of 
25%. If necessary you can push the joystick up or down to fine  
tune the full scale reading once)     
 

 

 

 

           

To return the main menu, push the Joystick left until 
the monitor scrolls Oxygen 0-25% on the top line. 

 
                                                           

NOTE:

 

To fine tune the reading to match the  

                                                                        reading on the oxygen monitor, use 
                                                                        the “Set 20mA Span” menu.  Push the  
                                                                        joystick up or down until both displays match

 

 
CAUTION: 

 

If the oxygen monitor is in the measuring mode and the ZERO push button is 

pressed an “Err” will be displayed on the remote display and no readings will be recorded.  

If the 

SPAN push button is pressed in the measuring mode, the remote display will automatically 
adjust to 25.0%. 

 You will need to repeat the above adjustment procedure again for the remote 

display to mirror the oxygen monitor.   

NEVER PUSH THE SPAN 

in the measuring mode.  

 

How to set the internal alarms on the Remote Display 

 
The Remote Display has two user selectable alarms with a separate horn tone to identify Alarm 1 
from Alarm 2.  Both are preset at the factory; AL1 = 19.5% and AL2 = 18%.  They can be 
changed in the field.  
 
To set Alarm 1, Press Alarm 1 for two seconds to enter the alarm change mode.  The display will 
show the current alarm setting and the right most decimal point will be lit.  Use the UP or 
DOWN buttons to increase or decrease the alarm value.  When done, push the Alarm 1 button 
for 2 seconds to get back to the monitoring mode.  
 
To set Alarm 2, Press the Alarm 2 button for two seconds to enter the alarm change mode.  The 
display will show the current alarm setting and the left most decimal point will be lit.  Use the 
UP or DOWN buttons to increase or decrease the alarm value.  When done, push the Alarm 2 
button for 2 seconds to get back to the monitoring mode.  

 

 

Summary of Contents for 99091

Page 1: ...itor s input power and mA output connection An 18 AWG three conductor shielded control and instrumentation cable Belden 8770 or equivalent is recommended for the connection The maximum permissible cable length is 250 feet You can simultaneously connect one remote display to one Air Check Oxygen monitor and one programmable logic controller PLC PureAire Monitoring Systems Inc 1140 Ensell Road Lake ...

Page 2: ...l from the case Mount the case to a wall or other flat surface There are 4ea through holes for fastening the case The case is designed to be flush mounted NOTE The front cover can be removed from the case by gently pulling on the plastic hinges they will stay connected to the front cover To connect the front cover just push the plastic hinge into the case ...

Page 3: ... used with our toxic gas monitors the polarity is set to Above Normally set at the factory 3 Range Auto Adjustment Switches These switches are used to adjust the zero and full scale reading on the Remote Display Alarm Indicator When sending a 4mA and 20mA Max distance 250 feet CAUTION Make all wire terminations with the power OFF Failure to do so can damage the oxygen monitor and remote display an...

Page 4: ...w to Setup and Operate the Remote Display CAUTION Entering the set up menus will disable the real time oxygen readings After wiring the remote display to your Oxygen monitor turn on the power to your oxygen monitor During the warm up the remote display will illuminate a yellow fault LED and the display will read FLt indicating that the oxygen monitor is in fault This will clear after the oxygen mo...

Page 5: ...NOTE To fine tune the reading to match the reading on the oxygen monitor use the Set 20mA Span menu Push the joystick up or down until both displays match CAUTION If the oxygen monitor is in the measuring mode and the ZERO push button is pressed an Err will be displayed on the remote display and no readings will be recorded If the SPAN push button is pressed in the measuring mode the remote displa...

Reviews: