How to set the zero and full scale range
This will need to be set up in the field to adjust to the length of cable between the oxygen
monitor and the remote display alarm indicator. To perform this in the field, first enter the
password on your oxygen monitor and select the “Set 4-20mA loop” menu
Push the joystick right to select “Set 4mA Zero” and then
push the joystick right one more
time to activate it.
This sends a 4mA signal to the remote display. On the back of the remote
display push the ZERO button for 3 seconds. You’ll see the display reset and automatically
adjust itself to the new zero. If necessary, you can push the joystick up or down to fine tune the
zero reading.
NOTE:
Both AL1 and AL2 LED’s will also be illuminated.
(To exit the Zero menu, push the Joystick left once)
Push the joystick down once to select “Set 20mA Span” and
then
push the joystick right one more time to activate it.
This
sends a 20mA signal to the remote display. On the back of the
remote display push the SPAN button for 3 seconds. You’ll see the
display reset and automatically adjust itself to the new span of
25%. If necessary you can push the joystick up or down to fine
tune the full scale reading once)
To return the main menu, push the Joystick left until
the monitor scrolls Oxygen 0-25% on the top line.
NOTE:
To fine tune the reading to match the
reading on the oxygen monitor, use
the “Set 20mA Span” menu. Push the
joystick up or down until both displays match
.
CAUTION:
If the oxygen monitor is in the measuring mode and the ZERO push button is
pressed an “Err” will be displayed on the remote display and no readings will be recorded.
If the
SPAN push button is pressed in the measuring mode, the remote display will automatically
adjust to 25.0%.
You will need to repeat the above adjustment procedure again for the remote
display to mirror the oxygen monitor.
NEVER PUSH THE SPAN
in the measuring mode.
How to set the internal alarms on the Remote Display
The Remote Display has two user selectable alarms with a separate horn tone to identify Alarm 1
from Alarm 2. Both are preset at the factory; AL1 = 19.5% and AL2 = 18%. They can be
changed in the field.
To set Alarm 1, Press Alarm 1 for two seconds to enter the alarm change mode. The display will
show the current alarm setting and the right most decimal point will be lit. Use the UP or
DOWN buttons to increase or decrease the alarm value. When done, push the Alarm 1 button
for 2 seconds to get back to the monitoring mode.
To set Alarm 2, Press the Alarm 2 button for two seconds to enter the alarm change mode. The
display will show the current alarm setting and the left most decimal point will be lit. Use the
UP or DOWN buttons to increase or decrease the alarm value. When done, push the Alarm 2
button for 2 seconds to get back to the monitoring mode.