signal from your oxygen monitor, pushing these switches will automatically adjust the
zero and full scale readings on the Remote Digital Display.
4.
Alarm Adjusting Buttons
– These buttons are used to increase, (UP) or decrease
(DOWN) the alarm settings.
5.
Horn On/Off Switch
– This switch is used to turn the internal audio horn on and off. In
the On position, any alarm will activate the audible horn. To disable the internal horn,
turn the switch to Off.
How to Setup and Operate the Remote Display
CAUTION:
Entering the set up menus will disable the real time oxygen readings
After wiring the remote display to your Oxygen monitor, turn on the power to your oxygen
monitor. During the warm up the remote display will illuminate a yellow fault LED and the
display will read FL
t
, (indicating that the oxygen monitor is in fault). This will clear after the
oxygen monitor completes the 4 minute warm-up.
1. Alarm
adjusting
switches
2. Alarm
Polarity
switches
3. Range auto
adjustment
switches
Do not remove
jumper unless
connecting to
PLC.
4. Alarm
adjusting
buttons
5. Horn On Off
Switch
During warm up
After warm up