PureAire Monitoring Systems, Inc.
41
8.
Place the retainer ring over the membrane and then place the box nut over the retainer ring. Turn
clockwise until hand tight; continue tightening until the box nut can no longer be moved by hand.
CAUTION: If the box nut is too loose, the sensor cell response may be affected.
BOX NUT
RETAINER
RING
Hold the sensor in your hand with the electrode down. Pour fresh electrolyte in until it reaches the
MAX. mark.
PRESSURE COMPENSATION
PORT
ELECTROLYTE
LEVEL
ENLARGED
VIEW
NOTE: MAX. level differs from sensor to sensor.
9.
Wipe the pressure compensation port and its surrounding area with a dry paper tissue.
PRESSURE COMPENSATION
PORT
PAPER TISSUE
NOTE: The
Methyl bromide
sensor cell uses an
8mm diffuser.
Please DO NOT
REMOVE THIS
DIFFUSER
41