background image

Rev: A1 

page

19

of

23

 

 

◆ 

Electromagnetic Compatibility Guide

 

Important information regarding Electro Magnetic Compatibility (EMC) 

 

With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, 
medical devices in use may be susceptible to electromagnetic interference from other devices. 
Electromagnetic interference may result in incorrect operation of the medical device and create a 
potentially unsafe situation.   
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to 
prevent unsafe product situation, the IEC60601-1-2 standard has been implemented. This standard 
defines the levels of immunity to electromagnetic interferences as well as maximum levels of 
electromagnetic emissions for medical devices.   
This medical device manufactured by pump conforms to this IEC60601-1-2:2007 standard for both 
immunity and emissions.   
Nevertheless, special precautions need to be observed:   

·

Do not use mobile (cellular) telephones and other devices, which generate strong electrical or 

electromagnetic fields, near the medical device. This may result in incorrect operation of the unit 
and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. 
Verify correct operation of the device in case the distance is shorter.   
Further documentation in accordance with IEC60601-1-2:2007 is available at pump at the address 
mentioned in this user manual. 
 

Guidance and manufacturer’s declaration Guidance and manufacturer’s declaration – 

electromagnetic emissions 

 

The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment 
specified below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure 
that it is used in such an environment.   

Emissions test 

 

Compliance 

 

Electromagnetic 

environment - guidance 

 

RF emissions   
CISPR 11   

Group 1   

The [EQUIPMENT or 
SYSTEM] uses RF energy 
only for its internal function. 
Therefore, its RF emissions 
are very low and are not 
likely to cause any 
interference in nearby 
electronic equipment.   

RF emissions   
CISPR 11   

Class B   

 

Harmonic emissions   
IEC 61000-3-2   

Not applicable   

 

Voltage fluctuations/flicker 
emissions  
IEC 61000-3-3   

Not applicable   

 

 

Summary of Contents for SF1402B

Page 1: ...Instruction Manual z Thanks for purchasing the Body Composition Analyzer and Scale manufactured by Bpump z Please read the instruction manual before usage z Please dispose the Body Composition Analyze...

Page 2: ...4 3 Display the Measurement Result 10 4 4 4 Data Transmission 11 5 Reference Standards 12 5 1 Weight Standard 12 5 2 Body Fat Ratio Standard 12 5 3 Water Content Standard 12 5 4 Visceral Fat Index Sta...

Page 3: ...ct to the device z Please carefully dealing with broken glass pieces if device broken in avoid of hurt z Do not stand on the edge of the devicein avoid of inaccurate measurement z Do not jump on the d...

Page 4: ...during usage because portable and mobile radio frequency communication equipments may affect the device performance do not close to the cell phones microwave ovens z Please refer to electromagnetic co...

Page 5: ...y SET Key Right Arrow use to select mode or increase numericalvalues Left Arrow use to select mode or decreasenumericalvalues Visceral fat index Physical Age Bluetooth Battery symbol Group Measurement...

Page 6: ...n soaked with water or neutral detergent then wipe the device with another dry soft cloth otherwise may cause infection z The deviceis not intended to use except for the measurement of physiological i...

Page 7: ...uest function to measure physiological indexes such as weight the body fat ratio visceral fat index and others z What is guest function Instead of setting personal data in advance user can also measur...

Page 8: ...Steps for Measurements 4 4 2 1 Whole physiological Indexes Measurement 1 Power On Off Turn on the device and wait for a while then 0 0 kg will show on the screen fig 2 2 Personal Data Setting Set the...

Page 9: ...4 2 2 Weight Measurement Only 1 Power On Off Power on the device and wait for a while 0 0 kg will appearon the screen fig 2 Note automaticpower on can be triggered in casea person thingupon weighting...

Page 10: ...for a while the weight value will display on the screen fig 13 3 Step off the platform after the measurement done 4 4 3 Display the Measurement Result 4 4 3 1 Physiological Indexes Display The device...

Page 11: ...ss SET button for confirmation The available units are metric and imperial such as cm kg ft lb fig15 fig 16 4 4 4 Data Transmission The bluetooth function turns on in accompany with the powering on of...

Page 12: ...otal amount of fluid in the body expressed as a percentage of total weight By weight the average human adult male is approximately 60 water By weight the average human adult female is approximately 50...

Page 13: ...umption 3 2kg 3 95kg 4 4kg 2 2kg 2 8kg 2 9kg 5 8 Basal Metabolism Age Male Female ideal body weight kcal Basal metabolism standard values kcal kg da y Basal metabolism amount kcal day ideal body weigh...

Page 14: ...urement Do not sway or leave during measurement 2 Do not take measurement barefoot Take measurement barefoot 3 The device is unable to detect the human impedance Step onto the platform with correct po...

Page 15: ...ulation Take measurement after blood circulation back to normal state Electrode is too cold Place the device in warm indoor place for a while Take measurement after electrode become warm The soles of...

Page 16: ...Display accuracy Weight 0 1kg Body Fat Rate 0 1 Water Content 0 1 Muscle Mass 0 1kg Bone Content 0 1kg Basal Metabolism 1Kcal BMI 0 1 Physical Age 1year Visceral Fat Index 1 SubregionalDeter mination...

Page 17: ...may cause undesired operation Note This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pro...

Page 18: ...evice needing clean wipe the device with wrung soft cloth which had been soaked in water or neutral detergent then wipe the device with another dry soft cloth 9 2 Storage Tips Please do not storage in...

Page 19: ...other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the unit and create a potentially unsafe situation Recommend...

Page 20: ...Not applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not applicable Mains power quality should...

Page 21: ...to 2 5 GHz Not applicable 3V m Portable and mobile RF communications equipment should be used no closer to any part of the EQUIPMENT or SYSTEM including cables than the recommended separation distance...

Page 22: ...ing to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2p 80 MHz...

Page 23: ...ce Registration xxxxxxxx After Sale Service Manufacturing Enterprises Shenzhen Pump Medical System Co Ltd ADD 2 F M 7 SinosteelBuilding Maqueling Estate Hi Tech Industrial Park Nanshan District Shenzh...

Reviews: