P842 Voltage Output
This parameter will determine the
range
of the
voltage output
. The voltage
output will
automatically
be
scaled
to represent the
Span
P106 where
0volts
represents 0% (
empty
) and maximum voltage
5
or
10 volts
,
depending on range selected, will represent 100% (
full
).
Option
Description
0 = 5 volts
Voltage output range will be 0 - 5volts.
1 = 10 volts
Voltage output range will be 0 - 10volts.
Additional Information
In the event of a
failsafe
condition
occurring
, the
voltage output
will fail
in accordance with the
Failsafe Mode P808
Compensation (CoP) Menu
P851 Measurement Offset
The value of this parameter is added to the measured distance, in
Measurement Units
(
P104
).
This Offset will be added to the level, as derieved from the transducer, and
will affect everything including the reading on the display, the relay
setpoints and the mA output.
P852 Temperature Source
Default = 1
This parameter determines the source of the temperature measurement. By
default
it is set to internal (transducer) (
P852=1
), which will automatically
detect the temperature from the transducer. If for any reason, no temperature
input is received, then the
Fixed Temp
value is used, as set by
P854
.
The temperature source can be specifically set as follows:
Option
Description
1 =
Internal (Default)
Always uses temperature reading from
transducer.
3 = Fixed
Always uses fixed temperature (P854)