Operating Manual - Error Messages and Erroneous Measurements
D909.131.00/02 en
71
Message
Measured quantities
Cause
Remedy
Reduced dose accu-
racy
Dose rate
Dose
Dose per pulse
Reduced dose accu-
racy, measuring error
≥ 5 %
Repeat the measurement.
Increase the dose rate on the
X-ray unit.
Reduce the distance.
Manually set the total filtration
in the
Measurement
menu.
Short exposure time
detected
Dose rate
Dose
Dose per pulse
The exposure time
is too short for the
required adapta-
tion of the measur-
ing range (refer also
to section
"Measurements with
Very Short Exposure
Times"
Repeat the measurement.
Low dose rate
detected
Dose rate
Dose
Dose per pulse
NOMEX is in a high
measuring range (refer
also to section
"Measurements with
Very Short Exposure
Times"
is too low for this mea-
suring range.
Repeat the measurement.
Dose rate too high
Dose measuring
range exceeded
Dose rate
Dose
Dose per pulse
Tube voltage
measuring range
exceeded
Tube voltage
Total filtration
Half value layer
The electronic mea-
suring system is out of
range.
Reduce the dose rate on the
X-ray unit.
Increase the distance.
Reduce the tube voltage or
tube current.
Dose rate too low
Dose rate
Dose
Dose per pulse
Insufficient kV signal
Tube voltage
Total filtration
Half value layer
The electronic mea-
suring system is out of
range.
Increase the dose rate on the
X-ray unit.
Reduce the distance.
Increase the tube voltage or
tube current.