NOMEX
Safety Information
D909.131.00/02 en
11
WARNING
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Exclusion of operation in the patient envi-
ronment:
Neither the product nor any peripheral devices
may be operated in the patient environment
(see
).
Exclusion of operation with patient contact:
The product is not intended for use in direct
contact with the patient. Neither the product nor
any peripheral device may have contact with
the patient.
Exclusion of operation as a controlling
instrument:
The product is only designed for use as a mea-
suring device. Measuring results must not be
used to control radiodiagnostic equipment or
radiotherapy units.
WARNING
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Use of peripheral devices:
Peripheral devices (PC, printer) may only be
connected if they meet the requirements of
standard UL 60950-1 or IEC 60950-1 or of stan-
dard IEC 61010-1.
Devices may be connected to other devices or
to parts of systems only if it has been ascer-
tained that this connection does not impair the
safety of the patient, the operator, or the envi-
ronment.
If the device specifications do not contain
information regarding the connecting of the
device to other equipment, you must consult
the manufacturer of the other equipment or an
expert about the effects of the connection on
the patient, operator, or environment. Always
observe standard IEC 61010-1.