DIADOS E
T11035
Service Manual
– Cleaning
D616.131.00/10 en
43
Service Manual
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
The following service tasks must only be per-
formed by specially trained staff.
Further service tasks may only be carried out by
PTW-Freiburg or by persons authorized by
PTW-Freiburg.
7 Cleaning
Device surface
Before cleaning or disinfecting the device sur-
face, disconnect the device from the power line.
Wipe the device clean with a moist cloth. Do not
let liquid enter the device. Do not apply spray to
the device.
Use a mild soap solution to clean the device.
Due to the device geometry, a thorough disinfec-
tion is not possible. The device cannot be steri-
lized.
WARNING
Liquids have entered the device.
Shock Hazard, Erroneous Measurement!
The device must be thoroughly inspected by
service personnel before being used again.
CAUTION
Disinfection of the surface with disinfectants on a
phenol base or peroxide compounds.
Equipment Damage!
Do
not
use disinfectants on a phenol base or
peroxide compounds to disinfect the surface.