DIADOS E
T11035
Operating Manual
– Description of the Device
D616.131.00/10 en
13
1.4 Keypad
Figure 4:
DIADOS E keypad
The second function of a key (indicated in italics) is
activated with the combination of ALT key + the re-
spective key.
Function
Effect
Comment
APPL.
selects the application
RAD, DENT, CT, MamM/M, MamM/R, MamW/A,
MamW/R, MamR/R, MamWAg
Only calibrated applications can be select-
ed.
FILTER
toggles between the calibration factors
in front of ↔ behind the phantom
combination of keys
ALT
+ APPL.
MODE
toggles between pulsed mode ↔ normal mode
in the pulsed mode, the unit is displayed in
reverse video
RANGE
toggles between the measuring ranges
Auto Range, Low, Medium and High
combination of keys
ALT
+ MODE
RESET
resets the display to zero
STA
RDY, TST
RDY
ZERO
zero adjustment
combination of keys
ALT
+ RESET
START/
STOP
manually starts/stops a dose measurement
ALT
activates the second function of a key
toggles between dose, dose rate and time in
normal mode and
between dose, dose/pulse and pulse in pulsed
mode
You can toggle during and after the meas-
urement.
ON/OFF
turns the device ON and OFF
push for 2 s
SETUP
brings up the Setup menu
combination of keys RESET + APPL.
,
used to navigate in the Setup menu
selects the sub-menu
ESC
used to exit the sub-menu/Setup menu