Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
8
D596.131.00/05 en
Safety Information
Strictly observe the following safety informa-
tion concerning the signal words DANGER and
WARNING. Failure to do so endangers the lives
of the patient, user, and other persons involved.
DANGER
Operation in areas where an explosion may
occur or in oxygen-enriched atmospheres.
Explosion Hazard!
Do not operate the product in areas of risk
where an
explosion
may occur. Explosion
hazards may be caused, for example, by the
use of combustible anesthetics, skin-cleansing
agents, and disinfectants.
Do not operate the product in
oxygen-
enriched atmospheres
. The atmosphere is
considered to be oxygen-enriched when more
than 25 % of oxygen or nitrous oxide is added
to the ambient air.
DANGER
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
– Before using the product, verify that it is in
correct working order and operating condi-
tion.
– Before putting the product into operation,
inspect all connection cables, connectors,
housing, and covers for signs of damage.
Replace damaged cables and connectors
immediately.
DANGER
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
– Switch on products on which moisture con-
densation has developed as a result of tem-
perature changes only after they are com-
pletely dry.
– Liquids must not enter the product. If liquids
have entered the product, dry the product
or wait for it to dry completely. The prod-
uct must be thoroughly inspected by service
personnel before being used again.