RT9200-01 Rev B                                                                                                                                                   Page 1 

The 

Prytime Medical™ STAAR (

Simulation Trainer for Arterial Access 

and REBOA) Assemble, Disassemble & Troubleshooting Instructions 

 

A.

 

Introduction 

Thank you for using the 

Prytime Medical™ STAAR (

Simulation Trainer for Arterial Access and REBOA) .  For additional 

information and product updates se

www.prytimemedical.com/product/staar

. 

B.

 

Assembly of the STAAR 

Open the case containing the STAAR components by 

turning the combination to ‘000’ and then

 pressing the zipper-

release button in towards the combination dials. 

 

1.

 

Unzip the case and remove the body model, hoses, tablet, sump reservoir case, pump case, pressure infusion 
bag, pressure transducer and all cables. 

2.

 

Fill the sump with water.  For each fill, add 4 drops of the included algaecide solution to the water.  

Note:

  The 

algaecide is effective for approximately one week.  Weekly changing of water with addition of solution is 
recommended. 

 

 

Summary of Contents for STAAR

Page 1: ...Assembly of the STAAR Open the case containing the STAAR components by turning the combination to 000 and then pressing the zipper release button in towards the combination dials 1 Unzip the case and...

Page 2: ...ith the intention that the sump and pump are to be placed on the floor and the simulator body is to be placed at table height If the pump and sump are placed even with the body the preset pressures wi...

Page 3: ...the body model Remove the guidewire and dilator 6 Attach the blood pressure monitoring cable and USB cable to the pump Attach the blood pressure monitoring transducer s extension line to the ER REBOA...

Page 4: ...automatically start when the pump is powered on If the application is open turn it off before proceeding to the next step by pressing the Close App button on the front of the tablet 9 Plug the tablet...

Page 5: ...ected to the tubing with the body model in both the front and back Ensure The ER REBOA Catheter s ART line stopcock is closed to the atmosphere 12 Once all potential leak points have been checked acti...

Page 6: ...tolic pressure of the model should be approximately 60 80 mmHg with the ideal pressure being 80mmHg The two sliding bars on the display screen labeled Intensity and Offset can be used to adjust this p...

Page 7: ...pressure in the bag Note The pressure infusion bag represents venous pressure so an approximate pressure of 10 15 mmHg should be estimated Do not over inflate the infusion bag 3 You will need to purge...

Page 8: ...Turn on the pump circulatory flow and while the pump is running disconnect the SUMP OUT yellow tube from the sump Note This will allow the pump to run dry while pumping water into the sump Allow the p...

Page 9: ...he USB cable is securely attached to the pump and the tablet Unplug and re plug as needed You can test and use the pump without the tablet by unlatching and opening the pump case and pressing the smal...

Page 10: ...the plug leaks water stop the pump and disengage the connection Inspect the o ring on the plug for damage Replace damaged o rings with new o rings found in the Repair Kit 7 Air is being pumped into th...