Procedures
Tsunami MP.11 5012/5054-SUI Installation and Management
Image File Download with the Bootloader
122
Image File Download with the Bootloader
The following procedures download an image file to the unit after the embedded software has been erased with
or when the embedded software cannot be started by the Bootloader. A new image file can be downloaded to the
unit with ScanTool, or the Command Line Interface through the unit’s serial port. In both cases, the file is transferred
through Ethernet with TFTP. Because the CLI serial port option requires a serial RS-232C cable, Proxim recommends the
ScanTool option.
Download with ScanTool
To download an image file with the ScanTool:
1. Set up the TFTP server as described in
2. Download the latest software from
.
3. Copy the latest software updates to your TFTP server’s root directory.
4. Run ScanTool on a computer that is connected to the same LAN subnet as the unit. ScanTool scans the subnet for
units and displays the found units in the main window. If in
, ScanTool does not find the device until the
unit Bootloader times out from its default operation to download an image. Click
Rescan
to re-scan the subnet and
update the display until the unit shows up in Bootloader mode.
5. Select the unit to which you want to download an image file and click Change.
6. Ensure that
IP Address Type Static
is selected and fill in the following details:
•
Password
•
IP Address and Subnet Mask of the unit.
•
TFTP Server IP Address
and, if necessary, the
Gateway IP Address
of the TFTP server.
•
Image File Name
of the file with the new image.
7. Click
OK
to start the file transfer.
The unit downloads the image file. The TFTP server program should show download activity after a few seconds. When
the download is complete, the LED pattern should return to
reboot
state. The unit is ready to start the embedded
software.
After a Forced Reload procedure, the unit returns to factory default settings and must be reconfigured. ScanTool can be
used to set the system name and IP address.
To access the 5012/5054-SUI see
Logging in to the Web Interface
Download with CLI
To use the CLI through the serial port of the unit, you need a standard serial connector and an ASCII terminal program
such as HyperTerminal. Proxim recommends you switch off the unit and the computer before connecting or disconnecting
the serial RS-232C cable.
To download an image file:
1. Set up the TFTP server as described in
2. Download the latest software from
.
3. Copy the latest software updates to your TFTP server’s root directory.
4. Use a straight-through serial cable to connect the unit’s serial port to your computer’s serial port.
5. Start the terminal program (such as HyperTerminal), set the following connection properties, and then connect:
•
Connect using: Com Port: <COM1, COM2, etc., depending on your computer>
•
Port Settings:
– Baud rate: 9600
– Data Bits: 8