Synchronous Cardioversion with LIFEPAK 5 Defibrillator
Warning
Use only the correct Protocol Systems cable with the LIFEPAK 5 Defibrillator as listed in the Protocol Systems
Products and Accessories guide when performing synchronous cardioversion. (This cable contains circuitry in
addition to wiring.) The use of any other cable will result in incorrect operation.
The Defibrillator Synchronization option is designed to operate only with the LIFEPAK 5 or
LIFEPAK 6s defibrillator. These instructions are not intended to replace existing hospital procedures
for cardiac electrical therapy and operation of the Physio-Control LIFEPAK 5 defibrillator. Follow all
safety standards and clinical protocols as defined by your institution.
Install the Interface Cable
1.
Before installing the LIFEPAK 5 Defibrillator Synchronization
Interface Cable (P/N 008-0136-XX) on the defibrillator,
examine the contacts on the left side of the LIFEPAK 5
defibrillator. Make sure the contacts are clean in order to allow
good signal transmission to the Propaq CS monitor.
2.
Slide the Interface Cable onto the left side of the
defibrillator as shown until it snaps in place.
3.
Connect the other end of the Interface Cable to
the DEFIB SYNCHRO connector on the monitor right
side panel.
Contact
points
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...