INS8800 Rev. D
13
Protimeter MMS2 Instruction Manual
4.5 Condensator
The
Condensator Mode
enables the user to assess the risk of condensation
occurring on surfaces or to confirm whether or not condensation is present on
a surface.
The MMS2 can be used as a
Condensato
r using two modes:
4.5a Surface Temperature Probe (Contact Based) - Selection
and Use
Navigate to
SELECT MODE -> SURFACE TEMPERATURE PROBE
and press
to
select the
Surface Temperature Probe
mode.
In this mode, the MMS2 measures the surface temperature using an external
Surface Temperature
probe inserted into socket
C
and making contact with the
surface to be assessed. In addition to the
Surface Temperature
probe, connect
a
Humidity
probe into socket
A
for the MMS2 to show the
Condensation
status.
TDIFF
is a useful function when investigating condensation, as it tells the user
how many degrees a surface temperature is above or below the prevailing dew
point temperature.
4.5b Surface Temperature IR (Non-Contact Based) - Selection
and Use
In this mode, the MMS2 measures surface temperature using IR technology
Connect a humidity probe into socket
A
.
Hold the
button to enable the
IR Thermometer
. Release the button and
press it again within 1 second to enable the
LASER
pointer. The
LASER
pointer
will indicate the area on the surface where the measurement is being taken
(see
Figure 5 on page 14
).
Summary of Contents for MMS2