9

After the first measurement is taken, a number will appear 
on the left side of the display along with the thickness 
reading on the right. The number on the left indicates the 
number of measurements that must be taken to complete 
the current series for the CAM average. Thus, each time a 
measurement is taken, the number on the left decreases by 
one.

The CAM indicator on the top of the display will flash on 
and off indicating that a CAM series is in progress. 
When the final measurement of the series is taken, the 
gauge will “beep” then the CAM average is display on the 
right and the average symbol “

<X>   “ 

on the left and the 

CAM indicator stops flashing.

To begin another series with the same number of 
readings, just begin measuring. This process can be 
repeated as many times as desired.

To begin another series using different number of 
readings, repeat the step 1 to step 3.

Turning off  CAM  Mode

1- Enter into CAM mode. “

CAM  MODE

” appears on 

    the display.

2- Press the 

   key.  “

CAM  OFF

” appears on the 

    display.

3- Press the 

ON

  key.  “

-- -- --

” will appear on the display 

    and the 

CAM 

indicator disappears.

Exit  CAM  Mode

1- Enter into CAM mode. “

CAM  MODE

” appears on 

    the display.

2- Press the            key.  “

CAM  OFF

” appears on the 

    display.

3- Press the            key again. “

CAM EXIT

” appears on

    the display.

Summary of Contents for PROTEK II PROTEK II SF

Page 1: ...Coating Thickness Instruments ProTek II S SERIES User s Guide ProTek II SF _____ ProTek II SN _____ ProTek II ST _____...

Page 2: ...t to fully utilize its capabilities 1 1 Applications MODEL Application ProTek II SF Measures the thickness of all non magnetic coatings and finishes on iron steel and other magnetic metal surfaces Pro...

Page 3: ...ries a Bare test plate for setting ZERO calibration ProTek II SF supplied with steel plate ProTek II SN supplied with aluminum plate ProTek II ST supplied with steel and aluminum plates b Two precisio...

Page 4: ...ode 3 Increases number of measurements in CAM mode 1 Enter CAM average mode 2 Decreases displayed value in calibration mode 3 Decreases number of measurements in CAM mode mils to micron to mils 3 1 Sp...

Page 5: ...ic shell and gently press the tip against the surface to be measured Use enough pressure to cause the outer shell to contact the measurement surface A thickness reading in mils or microns will be disp...

Page 6: ...er on by pressing the ON key 2 Holding the probe by its outer plastic shell place the tip on the uncoated metal part Gently push the probe shell until it contact the bare surface a If the display show...

Page 7: ...re metal part is rough or uneven b Otherwise press the ON key to continue Setting the Calibration 2nd Point 6 The display will flash W SHIM and continue to beep Place the plastic shim on the same bare...

Page 8: ...ment mode is a special feature which permits the user to take a series of consecutive measurements from 2 to 99 which are then averaged Enter CAM Mode 1 Press the key The display will show CAM MODE 2...

Page 9: ...ill beep then the CAM average is display on the right and the average symbol X on the left and the CAM indicator stops flashing To begin another series with the same number of readings just begin meas...

Page 10: ...a battery indicator appears on top of the display replace the battery Battery life is approximately 6 months when the ProTek II series gauges are used for about 80 readings per day BATTERY TYPE Any hi...

Reviews: