5.0  CALIBRATION

5.1  Introduction

The ProTek II series normally do not require calibration
before use. However, whenever the product or part to be
checked varies greatly from the supplied test plate, in
either shape or thickness, it is best to check or verify the
calibration using an uncoated sample of the actual product
or part and using one of the plastic calibration shims
provided.

Check the calibration as noted in section 5.2, and change 
the calibration, if necessary, following the procedure in
section 5.3. If a calibration change is necessary, continue 
to use the bare part instead of the supplied test plate.

5.2  Checking Calibration

1. Turn the power on by pressing the 

ON

 key.

2. Holding the probe by its outer plastic shell, place the tip 
    on the uncoated metal part. Gently push the probe shell
    until it contact the bare surface.

a. If the display shows near 

0.00,

 proceed to no. 3

b. If the display shows a reading more than 0.03 
    mils (0.8 um) on non-magnetic metals or more 
    than 0.01 mils(0.2 um) on magnetic metals, 
    proceed no further since the calibration must 
    be changed.

3. Place one of the plastic calibration shims on top of the 
    bare metal part, place the probe tip on top of the shim
    and metal part and measure the thickness of the shim.
    If the displayed thickness 

matches

 the thickness value 

    noted on the shim(within the accuracy limit) the 

  

    calibration is verified.

THE GAUGE IS NOW READY FOR USE.

 5.3  Changing Calibration

1. Turn the power on by pressing the 

ON

 key.

2. Press the

   key to start the calibration.

Summary of Contents for PROTEK II PROTEK II SF

Page 1: ...Coating Thickness Instruments ProTek II S SERIES User s Guide ProTek II SF _____ ProTek II SN _____ ProTek II ST _____...

Page 2: ...t to fully utilize its capabilities 1 1 Applications MODEL Application ProTek II SF Measures the thickness of all non magnetic coatings and finishes on iron steel and other magnetic metal surfaces Pro...

Page 3: ...ries a Bare test plate for setting ZERO calibration ProTek II SF supplied with steel plate ProTek II SN supplied with aluminum plate ProTek II ST supplied with steel and aluminum plates b Two precisio...

Page 4: ...ode 3 Increases number of measurements in CAM mode 1 Enter CAM average mode 2 Decreases displayed value in calibration mode 3 Decreases number of measurements in CAM mode mils to micron to mils 3 1 Sp...

Page 5: ...ic shell and gently press the tip against the surface to be measured Use enough pressure to cause the outer shell to contact the measurement surface A thickness reading in mils or microns will be disp...

Page 6: ...er on by pressing the ON key 2 Holding the probe by its outer plastic shell place the tip on the uncoated metal part Gently push the probe shell until it contact the bare surface a If the display show...

Page 7: ...re metal part is rough or uneven b Otherwise press the ON key to continue Setting the Calibration 2nd Point 6 The display will flash W SHIM and continue to beep Place the plastic shim on the same bare...

Page 8: ...ment mode is a special feature which permits the user to take a series of consecutive measurements from 2 to 99 which are then averaged Enter CAM Mode 1 Press the key The display will show CAM MODE 2...

Page 9: ...ill beep then the CAM average is display on the right and the average symbol X on the left and the CAM indicator stops flashing To begin another series with the same number of readings just begin meas...

Page 10: ...a battery indicator appears on top of the display replace the battery Battery life is approximately 6 months when the ProTek II series gauges are used for about 80 readings per day BATTERY TYPE Any hi...

Reviews: