background image

 

93-340-74 Issue 5                         CirrusPro Locator Operating Manual                             Page 11 of 12 

7.6 

Buzzer Enable Or Disable 

 

 

 

Enter the ‘Engineer menu’. (Section 7.5) 

 

Press the ‘increase sensitively’ button to enter menu. 
The LCD will show: 

 
 
 
 
 

 

SELECT

 either buzzer 

ON

 or 

OFF

 by toggling the ‘increase sensitively’ button. 

 

STORE

 the chosen option be pressing ‘decrease sensitively’ button that in turn exits back to 

the ‘Engineer menu’. 

 

 

8.0  Faults 

 

 

No Water 

The unit is out of water. 
Check  water  supply  and  refill.  Force  a  fill  by  pressing  the  hardware  reset  button.  A  fill  will 
occur within 2 minutes. Water should seen to be taken from the bottle. 
 
If water is not taken then the water fill path could be blocked or the valve has failed, in which 
case replace the cloud chamber. 
 

 

EEPROM Corrupt 

The  internal  program  memory  (eeprom)  does  internal  tests.  This  could  be  due  to  an  error 
during saving data. Re commissioning the unit and saving the settings should clear this error. 
If the error persists then replace the geartray assembly. 
  

 

Processor Fault 

This  is  indicated  if  a  problem  has  been  detected  with  micro  controller  or  the  Read  Only 
Memory  (ROM).  This  is  a  serious  failure  and  usually  requires  a  replacement  geartray 
assembly. 
 

 

Seal Fault 

The  chamber  is  tested  for  ‘leaks’  during  the  power  up  tests.  This  fault  does  not  necessarily 
stop the unit functioning, but it must be attended to before complete failure occurs. 
Check  LED/Photodiode  screws/seal.  Tighten  the  six  chamber  screws  slightly  and  check  the 
thermistor seal. 
If fault persists, replace cloud chamber. 

 

 

Vacuum Fault  

The  unit  has  detected  that  the  vacuum  is  insufficient  to  form  a  cloud  (<5.7psi).  Check 
connections to the pump, the pump operation and the pipe to sensor on the PCB and from the 
pump to the cloud chamber. 
If the fault remains, check that pump armature is parallel with the pump body, if not, adjust it 
down. Change the pump if required. 
 

 

Serial Fault  

 

 

 

Check the connection between the processor board (WEFA321) and the display control board 
(WEFA390). If the fault still occurs, replace either of the boards to determine which one  has 
the fault.

  

 
 
 
 
 

Summary of Contents for Cirrus Pro Locator

Page 1: ...Protec Fire Detection Export Ltd Protec House Churchill Way Nelson Lancashire BB9 6RT Telephone 44 0 1282 717171 Fax 44 0 1282 717273 CIRRUS PRO LOCATOR OPERATING MANUAL...

Page 2: ...THE CIRRUS PRO LOCATOR 5 5 0 OPERATING PROCEDURE 6 5 1 After Localizing the Fire Threat 8 6 0 TESTING AND DEMONSTRATING THE PRO LOCATOR 8 7 0 LCD MONITORING AND CONTROL 9 7 1 Standard Operation 9 7 2...

Page 3: ...n CAUTION Carefully read through this entire manual before using your new Cirrus Pro Locator Personal Safety Stay alert watch what you are doing and use common sense when operating the fire detector D...

Page 4: ...ge is 100V 240V AC 50 60Hz Only use approved charger Preparing your Cirrus PRO Locator Press the Power Button to turn unit on The power light will illuminate green The unit will initialise if the FAUL...

Page 5: ...graph on the display and four audible indicator tone speeds in conjunction with 10 selectable sensitivity settings As you get closer to the fire threat the percentage will increase and the audible ind...

Page 6: ...hich half of the area has a greater concentration of combustion 4 While slowly walking through the area watch bar graph level and audible indicator will increase or decrease according to how close you...

Page 7: ...with this divide and sample approach until you isolate the area down to a single room 8 When sampling in smaller areas there will be a higher concentration of combustion You may need to greatly decre...

Page 8: ...wire to simulate the overheating condition of a fire threat 1 Be sure you fully understand How to Use the Pro Locator Section 4 0 and the Operating Procedures Section 5 0 prior to any testing or demon...

Page 9: ...for the unit to describe the actual fault This is achieved by entering the engineer menu please refer to section 7 5 In the event of the water running out the LCD will display and the fault LED will i...

Page 10: ...urs 30mins 30min Count down from 5minutes NOTE if the batteries have NOT been FULLY charged these do NOT use the estimates Once all the battery cells are empty a 5 minutes count down will start and th...

Page 11: ...the error persists then replace the geartray assembly Processor Fault This is indicated if a problem has been detected with micro controller or the Read Only Memory ROM This is a serious failure and...

Page 12: ...it weekly 10 0 Technical Specifications Dimensions WxHxD 13 75 x 10 25 x 4 75 350 x 260 x 120mm Weight Standard 13 2 lbs 6 kg Extended battery life 15 4 lbs 7 kg Power 24v DC Two 12V 2 2 or 3 3 Ah SLA...

Reviews: