V4.0_2021-05-26
PROGNOST C Instructions for use
5073-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
21 von 30
6
Electrical data
6.1
Connection
NOTE
The PROGNOST C is in need of the following power supply
Power supply
220-240 Vac
Power frequency
50-60 Hz
Input current
2,5 A
It is intended that the central power supply of the PROGNOST C is always be connected to an supplied
X-ray generator or power box using a hard-wired connection. Select a connection that consider the
electrical specifications of the PROGNOST C like in the table above.
There is a central power supply connection on the radiographic table to which the X-ray floor stand
and the vertical X-ray system image receptor stand are also connected.
WARNING!
To lower the risk of an electrical shock, the device can only be run on a power
supply with a protective conductor.
6.2
Electromagnetic Compatibility (EMC) after EN 60601-1-2
CAUTION!
As a medical electrical device, the PROGNOST C is subject to special
precautionary measures regarding EMC and must be installed and
commissioned in accordance with the EMC information contained in the
accompanying documents.
CAUTION!
Portable RF communication devices (radio devices) should not be used closer
than 30 cm (12 inches) to the marked parts and cables of the PROGNOST C.
Failure to observe can lead to a reduction in the performance characteristics
of the device.
6.2.1
Guidelines and Manufacturers declaration – electromagnetic interference (non-life
supporting device)
The PROGNOST C is intended for use in the electromagnetic environment specified below. The
customer or the Operator of the radiographic system should assure that it is used in such an
environment.
Measurement
interference
Compliance
Electromagnetic Environment
RF emissions
CISPR 11
Group 1
The X-ray component uses RF energy exclusively for its
internal function. As a result, its RF emissions are very
low and it is unlikely that nearby electronic devices will
be influenced.