V1.0_2021-05-17
PROGNOST XP, XPE, XPE-Akku
Instructions for use
5038-0-0002
32 of 51
All mechanical parts of the PROGNOST XP-series, including accessories, may only be cleaned with a
wipe disinfection with suitable surface disinfectants (e.g., Melsept® SF, 15 min. application time at 2%
concentration). The information provided by the disinfectant manufacturer on concentrations and
application times must be observed.
WARNING!
No highly flammable disinfectants may be used! For safety reasons, spray
disinfection must not be carried out, as the spray mist could penetrate the
device and cause short circuits or corrosion.
If disinfectants are used that can form explosive gas mixtures, the device may
only be switched on again when the gas mixtures have evaporated!
5.4
Inspection and Maintenance
WARNING!
No maintenance or repair work may be performed while the PROGNOST XP
device is being used with a patient!
All maintenance and repair work may only be performed by personnel trained
or authorized by PROTEC.
5.4.1
Daily Monitoring before and during the Examination Operation
PROGNOST XP:
•
Locking of the wheels must be checked by operating the arresters
•
Locking and smooth running if the table must be checked by operating the brake bracket
PROGNOST XP & PROGNOST XPE-Akku:
•
Verify that the brake rollers are also braked if the arrester is in brake position
•
Check that the table top brake moves smoothly if the brake bracket is applied
•
Check the table top brake if the brake pedal is not actuated
•
Check the height of the table top. If there is a visible difference in height, perform a zero balancing
•
Perform a visual inspection of the power cable. If the power cord is damaged, stop powering the
table and notify the PROTEC service department to replace the cord
5.4.2
Regular Monitoring
5.4.2.1
Quality control by the operator
Quality checks for X-ray components must be performed at regular intervals in accordance with the
relevant national guidelines.
5.4.2.2
Safety-related Controls
In the interest of the patients, operators and external third parties, it is necessary that all checks
regarding operational safety and/or functionality of the unit are carried out regularly every 12 months
by the PROTEC customer service or a service authorized by PROTEC.
All components within the PROGNOST XP-series that may pose a risk due to wear and tear must be
inspected and, if necessary, replaced every 12 months by the PROTEC service department or a PROTEC
authorized service provider.
In the case that the scheduled checks are not carried out, PROTEC GmbH & Co. KG accepts no liability
for damage to the operator and third parties if and insofar as damage results from insufficient or non-
performed checks.