V4.0_2021-05-26
PROGNOST XS
Instructions for use
5034-0-0002
17 of 25
PROTEC Dealer:
Service Technician Name:
Service Technician No.:
Customer:
Customer No.:
Place of Installation:
PROGNOST XS No.: 7034-0-0000
Serial Number:
Annually for
systems with
less than 100
patients/day
Semi-annually
for systems with
more than 100
patients/day
Every 5 years for
systems with less
than 100
Patients/day
Every 4 years for
systems with
more than 100
patients/day
V
is
u
al
c
o
n
tr
o
l
Fu
n
ct
io
n
al
c
o
n
tr
o
l
Ex
ch
an
g
e
Ex
ch
an
g
e
1
Safety Inspection
OK
Not
OK
Note
1.1.
Check the parts of the X-ray system
table for damages
x
1.2.
All labels are still fixed and readable x
1.3.
Check the 70Nm tightening torque
of the guide rolls fixing screws
x
1.4.
Brake functionality of all 4 guide
rolls are working properly
x
1.5.
Check all fixing screws, tighten if
necessary (see installation manual)
x
1.6.
The table top has no relevant
damage. If necessary perform
control exposures
x