PRF-911 Concentric Ring
12
e. Select Ohms Mode by pressing the “Mode” button once of the PRS-801 or
PRS-812 Resistance Instrument.
1. Push Test Button
2. Resistance Should be < 20
Ω
(1) If proper connection is made to the BNC/Dual Banana adapter
and the center electrode is in contact with the metal test bed, the
resistance instrument will indicate low resistance.
(2) If proper continuity indication confirms positive contact of the
center electrode to the metal plate, proceed with testing the outer
electrode continuity, as described in point f.
(3) If no continuity indication appears, change the test lead
connection to the other BNC/Dual Banana adapter receptacle
and repeat the procedure.
(4) If no continuity indication appears, clean the electrode surface with 70%
solution of IPA, test the PRS-801 or PRS-812 per its operating instructions
and repeat the procedure. (i.e. Install the supplied shunt and test the low
range. Check the cable & connections as well.)
f. Remove the PRS-801 or PRS-812 test lead from the BNC/Dual Banana adapter
and install it in the BLACK OUTER RING electrode receptacle located on the
side of the fixture housing.
Figure 9: Confirming Electrode Contact to Test Bed