Pronk Technologies
2017 07-18
www.pronktech.com
800-541-9802
Page 20
©2015
Zero Stopwatch Time
‘W’
Poll Stopwatch Time
‘w’
Returns the stopwatch time in seconds
Diagnostic Data On
‘G’
1.4 and above. Default is on
Diagnostic Data off
‘g’
1.4 and above
Polls and Replies for Data mode (SW 1.5 and higher)
Function
Poll
Reply
Poll Pressure in
PSI * 100
‘p’
‘p,’yyyy,xxxx<cr><lf> Y is current pressure, X is max pressure ,
positive or negative
Poll Pressure in
mmHg
‘m’
‘m,’yyyy,xxxx<cr><lf> Y is current pressure, X is max pressure ,
positive or negative
Poll Volume in ul
‘v’
‘v,’xxxx<cr><lf> X is volume in ul.
Returns ‘NOT PRIMED’ or ‘NOT STARTED’ when appropriate.
Poll Rate in 0.1
ml/hr
‘r’
‘r,’xxxx<cr><lf> X is rate on the screen in 0.1 ml/hr.
Returns ‘NOT PRIMED’ when appropriate.
Poll YSI temp in
0.1C
‘y’
‘y,’xxxx<cr><lf> X is YSI temp in 0.1C
Commands and reports for Trumpet mode (SW 1.5 and higher)
Function
Activated
By
Reports
Turn on trumpet
curves at 10 sec
resolution
‘h’
reports “T,”XXX”V,YYY<cr><lf> where xxx is total reading
time in ms and yyy is total reading volume in ul. Trumpet curve
data does not start until the unit is primed, reset, and the flow
begins.
trumpet at 5 sec
‘i’
reports “T,”XXX”V,YYY<cr><lf> where xxx is total reading
time in ms and yyy is total reading volume in ul. Trumpet curve
data does not start until the unit is primed, reset, and the flow
begins.
trumpet at 1 sec
‘j’
reports “T,”XXX”V,YYY<cr><lf> where xxx is total reading
time in ms and yyy is total reading volume in ul. Trumpet curve
data does not start until the unit is primed, reset, and the flow
begins.
trumpet at 200
ms
‘k’
reports “T,”XXX”V,YYY<cr><lf> where xxx is total reading
time in ms and yyy is total reading volume in ul. Trumpet curve
data does not start until the unit is primed, reset, and the flow
begins.
pressure curve at
200ms
‘n’
Reports “T,” XXX “P,” YYY “X”,ZZZ<cr><lf> where xxx is
time since curve start in ms and yyy is current pressure, and zzz is
the maximum pressure. Pressure and Max pressure are always
in 0.01PSI. Pressure curves begin immediately, regardless of
prime status and continues to run until “q” is sent or a different
trumpet selection is made.
All curves off
‘q’
Turns pressure and trumpet curves off