3.
Press the float to 40 l/h using the flow sensor
ð
The error message should disappear.
4.
Hold the flow sensor in this position and tighten the mounting
clip
5.
Then re-adjust the flow required using the ball valve
6.
Acknowledge any error message that occurs
7.
Reset any possible consequences of this in the overall instal‐
lation
8.
Check the threaded connector for leak-tightness
7.3 Calibration
You have to regularly calibrate the sensor during
operation. That means: 24 hours after initial cali‐
bration and then weekly thereafter.
Please take into account any deviating national
regulations.
WARNING!
Danger from hazardous substances!
Possible consequence: Fatal or very serious inju‐
ries.
Please ensure when handling hazardous sub‐
stances that you have read the latest safety data
sheets provided by the manufacture of the haz‐
ardous substance. The actions required are
described in the safety data sheet. Check the
safety data sheet regularly and replace, if neces‐
sary, as the hazard potential of a substance can be
re-evaluated at any time based on new findings.
The system operator is responsible for ensuring
that these safety data sheets are available and that
they are kept up to date, as well as for producing
an associated hazard assessment for the worksta‐
tions affected.