03.2 Device’s preparation
On receipt of the product it is advised:
·
Remove the packaging and distribute the material such that all components are visible.
·
Verify that all components / parts in the attached list are present.
The appliance must be checked before each use in order to detect any malfunction and / or damage
caused by transport and / or storage. In particular, check:
·
General functionality of the device.
·
Cleaning the device (remember that lack of cleanliness can cause the risk of cross-infection).
·
Absence of cuts, holes, rips in the structure, including belts.
·
Correct fixing of all nuts, bolts and nuts.
·
Correct fixing of the belts.
·
Correct attachment of the belts.
·
State of use (moving parts, wheels, belts, covers)
·
Integrity of belts and covers.
·
Integrity of the components.
·
Integrity of the handles.
·
Lubrication of moving parts.
·
State of use of the wheels and braking system.
·
Operation of springs.
·
The emergency vehicle is equipped with a fastening system dedicated to the Promeba
product.
·
There are safety belts for the immobilization of the patient and they are intact and operating.
·
The welds are intact, without cracks or breaks.
·
No tube or sheet metal present curves or cracks
If the above conditions are met, the device can be considered ready for use; Otherwise, you must im-
mediately remove the device from the service and contact the manufacturer.
03
INSTALATION
Summary of Contents for PC-610/7
Page 2: ......
Page 8: ......
Page 18: ......
Page 20: ...USER MANUAL STRETCHER PC 610 7 Review 12 19 20 04 1 Elements PC 610 7 04 OPERATING 1 12 14...
Page 21: ...USER MANUAL STRETCHER PC 610 7 Review 12 19 21 04 OPERATING 13 04 1 Elements PC 610 7 2 8 4...
Page 24: ......
Page 26: ......
Page 56: ......