USER GUIDE //
CHAIR PS-251 + POWER TRACK PA-260 + LOCKING SYSTEM PA-281
Review 04/20
28
04
OPERATING
04.5 Operation of up and down stairs of the track system
Starting from the transport position and always with a minimum of two sanitary operators:
UP
· Attach the chair to the track as explained in section 04.1.8
· Extend the telescopic handles of the chair as explained in section 04.1.3
· Fold the backrest handles as explained in section 04.1.7
· Extend the footrest once the patient is sitting as explained in section 04.1.4
· Fasten the patient with the two belts supplied as explained in section 04.1.2
· Fix the patient’s head using the immobilizer located on the track (A)
· Set the track to “stand by” by pressing the ON / OFF button on the control panel (B)
· Approach the chair to the base of the ladder and gently recline the track drive above the first steps of
the ladder.
· The second operator must extend the telescopic handles before starting the ride and accompany at
all times during the course (2)
· Press the “UP” button to go up the stairs (3)
· Once the section of the stairs is finished, it is very important to fold all the necessary elements during
the plane transfer (4)
DOWN
· The operator in front of the patient should place the chair at the beginning of the ladder, should assist
the inclination of the track (3)
· Press the “DOWN” button to go down the stairs (3)
THE PATIENT SHOULD
STAY FIXED
AT ALL TIMES WHEN THE LIFT UP / DOWN IS OPERATED.
NOT PERMITTED IN ANY CASE MOBILIZE PATIENTS
WITHOUT
COMPLYING WITH ALL THE SAFETY MEASU-
RES ESTABLISHED.