7. RS232 Interface
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
page 23
7.9
Digits
command:
digits [digits]
without parameter:
shows the number of hidden decimal places
with parameter:
hides the give number of decimal places
parameter:
0 digits 4
example:
>
dig
0 decimal places blinded out
>
>
dig 3
3 decimal places blinded out
>
7.10
Status or FM Status
command:
status
fmstatus
without parameter:
prints out a list of the actual settings
example:
>
STATUS
FM3002 status
-------------
serial no. is 0801770101
filter is 1
integration time is 1000 ms
unit is µT
0 decimal places blinded out
sound is on
keys are unlocked
offset = -76755
slope = 23702056
factory calibration active
>
7.11
Default
command:
default
without parameter:
resets the instrument to factory settings (exemption: settings for
slope and offset are conserved)
example:
>
default
factory settings restored
>