Owner’s Manual
Following the cleaning instructions after each use will
prevent any remaining medication in the bottle from
drying and adhering to the mesh cap and the device
not nebulizing effectively.
Before storing or carrying, make sure to clean the
nebulizer instantly with distilled water after each use.
Step 1: Dismantle the nebulizer
Remove the mouthpiece and then remove the
medication chamber from the nebulizer.
See Figure 20.
ATTENTION:
Clean out the remaining
medication after each use.
Step 2: Clean the remaining
medication
Open the cover of the medication
chamber and discard the remaining
medication. Pour a small amount of
distilled water into the medication
chamber and close the cover. Turn
on the device to nebulize
the distilled water for 1–2 minutes
to remove residual medication
from the mesh.
Cleaning the Nebulizer
Figure 20