14 PRODIGIT
ˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍˍ
CURR
Purpose :
Constant load current setting.
Command syntax :
CURR:[:TRIG]{SP}{NR2}{;NL}
Description :
This command is used to set the load current level of 3320 series Electronic load module.
Note
:
1.The load current data must include decimal point, otherwise this command is unable to excute.
2.Unit for load current is ampere.
3.Please make sure the SPEC command and range I / II command before excute the load current
setting command.
4.If the programming load current level over the maximum specification of 3320 series load module,
the full scale current will be sent to the load module.
5.The most effective load current level can be set is the fifth digit after the decimal point.
Example :
1.CURR:TRIG 10.20
2.CURR:3.0
3.CURR:5.123
LOAD
Purpose :
Turn the Electronic load module input ON or OFF.
Command syntax :
LOAD[:TRIG]{SP}{ON|OFF}{;NL}
Description :
This command sets the Electronic load to sink current from DC power source.
LOAD ON:Electronic load is ready to sink current from DC power source.
LOAD OFF:Electronic load can not sink current from DC power source.
Example :
a.LOAD ON
b.LOAD:TRIG ON
c.LOAD:TRIG OFF