27
M-43-25
LUF-Series Lufran Manual
Revision - Date 03-01-04-18 Standard outlet, Standard Interface
Engineering Information (Continued):
Operator Interface Panel (OIP) (Continued):
Flow Calibration Menu:
The flow calibration menu is accessed through the secure menu only. Refer to the engineering
information section of this manual for the flow meter calibration procedure.
Figure 17: Flow Calibration Menu
SLC Setup Information
Panel
Description
HIGH FLOW
Manually-entered flow rate used as the high-end calibration point,
taken from the display on the flow meter itself.
HIGH RAW
Manually entered raw data value that corresponds to the high flow
value, taken from the RAW DATA display on this menu.
SLOPE CALC.
Calculated value from the HIGH FLOW, HIGH RAW, LOW FLOW
and LOW RAW fields.
RAW DATA
Information from the flow meter.
LOW FLOW
Manually-entered flow rate used as the low-end calibration point,
taken from the display on the flow meter itself.
LOW RAW
Manually entered raw data value that corresponds to the low flow
value, taken from the RAW DATA display on this menu.
OFFSET CALC.
Calculated value from the HIGH FLOW, HIGH RAW, LOW FLOW
and LOW RAW fields.
FLOW CALC.
Calculated value from the SLOPE and OFFSET fields. When the
flow meter is calibrated, this value should match the display on the
flow meter.
SLOPE
Manually-entered field taken from the SLOPE CALC. field.
OFFSET
Manually-entered field taken from the OFFSET CALC. field.
BACK
This button will return to the previous screen.