When selected the Controller Output will be available on the second
analog output channel of the sensor
The display will then show the following control parameters:
CON TARG: 0.00%
This is the set point target. Its default value is 0.000
Minimum value is 0.00
Maximum value is 100.00
CON PROP: 1.00
This is the Gain of the control output. It is the ‘step’ change in the output that is made to correct the
difference between the measured variable and the set point
Gain = Output/Input
Increasing the value will cause the output to move in a larger step.
Minimum value is 0.0
Maximum value is 99.99
CON INT: 1.00 sec
This is the automatic reset that moves the output whenever there is an error between the process variable
and the set point.
The value is measured in SECONDS. This is the time it takes the reset (or integral) element to repeat the
action of the proportional element.
The lower the value the faster the proportional action takes place, the higher the value the less
frequently it takes place.
Default value is 1.00
Minimum Value is 0.00
Maximum Value is 99.99
CON DEAD: 0.00%
This is the DEAD-Band of the control system. When the measured variable is inside the Dead-Band the
integral reset will stop.
The Dead-Band is a value +/- either side of the set point.
Default value is 0.00
Minimum value is 0.00
Maximum value is 99.99