502 010 07
Manual
Flow Transmitter, PD 340
39/59
6.2
Factory settings
If specific functions are not requested at time of order, the transmitter will be delivered with
the following factory settings. The digits 7 and 8 are set to zero. In the rest of the tables in
the manual, these two digits are left out, as they are not accessible from the PD 210 Display.
PD 210 address
Variable name
C 25
C 38
C 51
C 63
C 76
C 102
E1 Meter Size
8.0
20.0
40.0
80.0
120.0
200.0
E5 Scale
Standard
Extended
.000010
8.00000
.000010
20.0000
.000100
40.0000
.000100
80.0000
.000100
120.000
.000100
200.000
E6 Display resolution
333333
333333
222222
222222
222222
222222
E7 Function selector Standard
Extended
411100
211100
411100
211100
411100
211100
411100
211100
411100
211100
411100
211100
E8 Function selector
302411
302411
302411
302411
302411
302411
Note that the E5 scale value for standard version in the table is keyed-in in m³ pr. pulse, due
to the corresponding E1 setting in m³, whereas the resulting output below is expressed in
litres pr. pulse.
The above factory settings result in the following outputs:
Standard version
:
Output2:
1 litre/pulse (C 25 and C 38)
10 litres/pulse (C 51, C 63, C 76 and C 102)
Output3:
0.01 litre/pulse (C 25 and C 38)
0.1 litre/pulse (C 51, C 63, C 76 and C 102)
Extended version
:
Output2:
1 litre/pulse (C25 and C38)
10 litres/pulse (C51, C63, C76 and C102)
Output3:
20 mA at max. flow rate