6.2 Calibration standard
Always protect the calibration standard from dust, moisture and other environmental
factors. It is recommended to keep it stored in the ZRM 6006 case. The standard
values measured in the factory as well as the serial number of the corresponding
instrument can be found on the sticker on the inside of the standard.
It is valid for 2 years and should be replaced or recalibrated by Zehntner after this
time.
The working standard delivered with the ZRM 6006 is not changeable and is
valid only for the delivered instrument.
6.3 Cleaning of the calibration standard
If the calibration standard is soiled or covered by dust, it can be cleaned carefully
using window cleaner and a soft tissue.
A damaged or a polluted calibration standard may cause incorrect calibration
and therefore incorrect measuring results.
Label:
•
serial number
•
calibration values
•
validity
snap lock
calibration standard
release