Seite 34
startup reminding you to lubricate the guide rods. Carry out the lubrication as
described in chapter 11.3 on page 37.
To switch off the reminder
until the next interval, choose “Do not show this message
again” and close with OK.
7.6 Emergency Stop function
Pressing the Emergency Stop button (11) immediately stops the motor. The power
supply of the motor is interrupted so that the drawing unit (23) can be moved by
hand.
The touchscreen and control unit of the
Proceq ZAA 2600 remain active since
only the power supply is interrupted.
To unlock the Emergency Stop button
(11), turn it counterclockwise.
The following message will be
displayed after unlocking the
Emergency Stop:
The Zero position must be reset. Move
the drawing unit (23) to the desired
starting position and press the „Zero”
button.