6.3 Carry out crockmeter tests
Never touch any moving parts during operation.
The side panels must always be mounted during operation.
Do not lean over the unit during operation.
Risk of injury: Never hold the sample by hand during operation. It has to be
fixed with an appropriate means, e.g. double sided adhesive tape.
Crockmeter tests are carried out in machine mode
„Crock/Scrub“. Press
menu on the touchscreen -> Setup -> Machine mode, choose
„Crock/Scrub“,
the sign
appears in the display.
•
Place the device on a solid, even surface.
•
Level the apparatus using the included spirit level. To do so, place the spirit
level on the sample tray and adjust the levelling feet (8) until the Proceq ZAA
2600 is levelled perfectly. The glass plate must be firmly in place.
•
Place the sample.
-
Use the glass plate with the glass side facing up for very thin samples.
-
Use the glass plate with the printing blanket facing up for thicker samples.
Fix the samples with a double sided adhesive tape on the glass plate in
order to get homogeneous friction lanes.
•
Move the tool holder (18) either with
or manually to the back position of
the sample.
•
Mount the chosen friction finger set (see chapter 6.2